The benchmark for emc
.
Operating-
manual
CDN 118A Series
CDN 118A-C4-4-1
CDN 118A-C6-4-1
Coupling network for surge and
telecom-surge to signal and datalines
The coupling network of the CDN 118A series are designed for
coupling surge and telecom-surge pulses to symmetrical signal
or datalines.
The coupling network expands the range of application of the
impulse generator of the NSG 3040A and NSG 3060A series,
for testing signal and datalines.
Main characteristics are the advantages for handling and the
easy using with banana plugs.
Version:
1.00 / 30.04.2019
Replaces: No previous version
Filename: UserManual-CDN 118A-series-E-V1.00.doc
Printdate: 30.04.19
EC 61000-4-5 Ed3.0
EN 61000-4-5
ITU-T K.44

Advertisement

Table of Contents
loading

Summary of Contents for Ametek CDN 118A Series

  • Page 1 CDN 118A-C6-4-1 Coupling network for surge and telecom-surge to signal and datalines The coupling network of the CDN 118A series are designed for coupling surge and telecom-surge pulses to symmetrical signal or datalines. The coupling network expands the range of application of the impulse generator of the NSG 3040A and NSG 3060A series, •...
  • Page 2 4153 Reinach BL1 Switzerland Phone: +41 61 204 41 11 Fax: +41 61 204 41 00 URL: http://www.emtest.com Copyright © 2019 AMETEK CTS GmbH. All right reserved. Information in earlier versions. Specifications subject to change Operating manual V 1.00 2 / 26...
  • Page 3: Table Of Contents

    Testing and danger ..........................6 2.2.1. Coupling networks ..........................6 2.2.2. Danger from EUT ..........................6 Standards for testing with coupling networks of CDN 118A series ......7 3.1. Device models ............................ 8 3.2. Impulse generators ..........................8 Device functions and operating ..................9 4.1.
  • Page 4: General

    1.1. Intended use Coupling decoupling networks of the CDN 118A series are used for simulating conducted electromagnetic inter- ference effects for immunity testing to international, national, and manufacturers’ standards. The CDN’s are designed for full compliance conducted electromagnetic compatibility (EMC) test requirements.
  • Page 5: Safety

    AMETEK CTS CDN 118A series Safety 2.1. Safety aspects Read the following operation manual carefully. Pay special attention to both safety and operation details!!! Observe all of these precautions to ensure your personal safety and to prevent damage to the test equipment.
  • Page 6: Testing And Danger

    AMETEK CTS CDN 118A series 2.2. Testing and danger All tests offered by the High Voltage or EMC generators are immunity tests on electronic equipment or devices. These tests are potentially dangerous for the operator. Therefore, it is the responsibility of the user to avoid criti- cal failures and risks to the environment and operator.
  • Page 7: Standards For Testing With Coupling Networks Of Cdn 118A Series

    AMETEK CTS CDN 118A series Standards for testing with coupling networks of CDN 118A series The coupling network CDN118A series are suitable for tests according the following standards: - IEC 61000-4-5 - Surge 1.2/50s as per figure 10 for unshielded symmetrical interconnection lines...
  • Page 8: Device Models

    4 x 25 Ω CDN 118A-C6-4-1 4 wires 50 VDC 6,6 kV 1.2/50us 3.2. Impulse generators The CDN 118A series can be used for the following surge generators: Device previous name Pulseform Remark generator NSG 3040A All models up to 4800V as per IEC 61000-4-5 1.2/50s - 8/20s...
  • Page 9: Device Functions And Operating

    AMETEK CTS CDN 118A series Device functions and operating The difference between the various models is: Number of data lines (CDN 118A Cx-4-1 series with 4 lines) nominal voltage and current level of the signal- and data lines various test voltage level of 4.8 kV and 6.6 kV 4.1.
  • Page 10: Test Setup For I/O And Datalines

    AMETEK CTS CDN 118A series 4.3. Test Setup for I/O and datalines Arrangement: The CDN 118A series is placed on the top or on the side of the surge generator. Connections frontside Couplings: All couplings for surge pulses are realized with short circuit connectors.
  • Page 11: Test Setup For 1.2 Μs / 50 Μs Impulses

    AMETEK CTS CDN 118A series 4.4. Test setup for 1.2 µs / 50 µs impulses The energy transfer from the surge generator to the EUT is considered to be a constant which independs from the number of lines in the cable, equivalent to a coupling impedance of about 40 Ω.
  • Page 12: Test Setup For 10Μs / 700Μs Impulses

    AMETEK CTS CDN 118A series 4.5. Test setup for 10µs / 700µs impulses Due to the nature of the wiring used for unshielded outdoor symmetrical communication lines (twisted pairs), the coupling is always in common mode. The coupling decoupling schematic is shown in Figure A.4. (IEC 61000-4-5 Ed 3.0)
  • Page 13: Diagram Cdn 118A Series

    AMETEK CTS CDN 118A series 4.5.2. Diagram CDN 118A-series Diagram CDN 118A-Cx-4-1 Operating manual V 1.00 13 / 26...
  • Page 14: Technical Data

    AMETEK CTS CDN 118A series Technical data 5.1. CDN 118A series CDN 118A-C4-4-1 CDN 118A-C6-4-1 Testlevel Max. test voltage 4800V 1,2/50s or 10/700s 6600V 1,2/50s or 10/700s Number of lines Decoupling L = 20mH L = 20mH Residual voltage (decoupled side) 1.2 µs / 50 µs...
  • Page 15: Weight And Dimension

    AMETEK CTS CDN 118A series 5.2. Weight and dimension Model Signal lines Weight [kg] Dimension Housing L x W x H [mm] CDN 118A-C4-4-1 4 wires 11.2 kg 448 x 500 x 154 CDN 118A-C6-4-1 4 wires 11.2 kg 448 x 500 x 154 Housing A =>...
  • Page 16: Maintenance

    AMETEK CTS CDN 118A series Maintenance 6.1. General The coupling network is built with passive components only. The coupling network of the CDN 118A series is absolutely maintenance free. 6.2. Calibration and Verification 6.2.1. Factory calibration Every AMETEK CTS generator is entirely checked and calibrated as per international standard regulations before delivery.
  • Page 17: Maintenance, Adjustments, Replacement Of Parts

    AMETEK CTS CDN 118A series 6.3. Maintenance, Adjustments, Replacement of Parts Electrical maintenance must only be performed by qualified service technicians... ATTENTION The generators do not contain any parts or components requiring special maintenance. Electrical maintenance must only be performed by experienced and specially trained technicians. Generally, standard maintenance requires only the periodic cleaning of the instrument, verification and calibration of certain parameters.
  • Page 18: Calibration Procedure As Per Iec 61000-4-5 Ed3

    AMETEK CTS CDN 118A series Calibration procedure as per IEC 61000-4-5 Ed3 Calibration Surge impulse (1.2/50 μs – 8/20 μs) 7.1. 7.1.1. General Calibration measurements shall be performed as indicated in Table 9 of IEC 61000-4-5 Ed3.0 at the CDN rated impulse voltage.
  • Page 19: Surge Waveform Definition On Cdn For Symmetrical Interconnection Lines

    AMETEK CTS CDN 118A series The intention of this calibration process is to check the proper function of the components, the saturation of de- coupling chokes, the decoupling effect of the decoupling part, the current capability and the coupling effect of the coupling network part.
  • Page 20: Calibration Telecom Surge Impulse (10/700 Μs - 5/320 Μs)

    AMETEK CTS CDN 118A series Calibration Telecom Surge impulse (10/700 μs – 5/320 μs) 7.2. 7.2.1. General Due to the nature of the wiring used for unshielded outdoor symmetrical communication lines (twisted pairs), the coupling is always in common mode.
  • Page 21: Telecom Surge Waveform Definition

    AMETEK CTS CDN 118A series The intention of the calibration process is to check the proper function of the components, the saturation of de- coupling chokes, the decoupling effect of the DN part, the current capability and the coupling effect of the CN part.
  • Page 22: Delivery Groups

    AMETEK CTS CDN 118A series Delivery Groups Identical accessory parts are delivered only once if several devices are ordered. The delivered packing list is in each case valid for the delivery. 8.1. Delivery groups CDN 118A-Cx-4-1 • Base equipment CDN 118A-C4-4-1 / CDN 118A-C6-4-1 •...
  • Page 23: Appendix

    AMETEK CTS CDN 118A series Appendix 9.1. Declaration of conformity Manufacturer: AMETEK CTS GmbH Address: Sternenhofstr. 15 CH 4153 Reinach BL1 Switzerland declare, that under is sole responsibility, the product’s listed below, including all their options, are conformity with the applicable CE directives listed below using the relevant section of the following EC standards and other normative documents.
  • Page 25: Typical Waves

    AMETEK CTS CDN 118A series 9.2. Typical waves 9.2.1. Voltage waveform Surge 1.2/50 μs Telecom Surge 10/700 μs 5.0 kV 5.0 kV Nominal Measured Nominal Measured U peak 5.0 kV ± 10% 4.86 kV U peak 5.0 kV ± 10% 4.84 kV...
  • Page 26: Residual Voltage Waveform

    AMETEK CTS CDN 118A series 9.2.3. Residual voltage waveform Surge 1.2/50 μs Telecom Surge 10/700 μs 5.0 kV 5.0 kV Residual voltage: < 80 V Residual voltage: < 83 V 9.3. Transferfunktion The figure below illustrates the transferfunction of a typical symmetrical lines L1-L2 & L3-L4.