PRECAUTIONS
•
Read your C
®
ontour
device package insert and all instructional materials provided in
your meter kit before testing. Please follow all instructions for use
and care exactly as described to help avoid inaccurate results.
• Examine product for missing, damaged, or broken parts. If the test
strip bottle is open inside a new box of strips, do not use those
strips. Contact Customer Service. See the back cover for contact
information.
• The C
®
n
meter only works with C
ontour
ext
strips and C
n
®
ontour
ext
• Always keep the C
ontour
Tightly close the bottle immediately after removing a test strip.
The bottle is designed to keep the test strips dry. Avoid exposing
meter and test strips to excessive humidity, heat, cold, dust, or
dirt. Exposure to room humidity by leaving the bottle open or not
storing the strips in their original bottle can damage your test
strips. This could lead to inaccurate results. Do not use a test strip
that appears damaged or has been used.
• Check the expiration dates on your test strips and control solution.
Do not use the test strips or control solution if the expiration date
printed on the bottle label and carton has passed. This can cause
inaccurate results. For the control solution, do not use if it has
been six months since you first opened the bottle. It will help to
write the six month discard date on the control solution label.
• If your control solution test result is out of range, contact Customer
Service. See the back cover for contact information. Do not use
the meter for blood glucose testing until you resolve this issue.
• The meter has been designed to give accurate results at
temperatures between 5°C and 45°C. If you are outside this range,
you should not test. Whenever the meter is moved from one
location to another, allow approximately 20 minutes for the meter
to adjust to the temperature of the new location before performing
a blood glucose test.
• Do not perform a blood glucose test when the C
meter is connected to a computer.
iii
®
85476920_CntrNXT_UG_EN.indd 3-4
n
meter user guide, the lancing
ext
®
n
test
ontour
ext
control solution.
n
test strips in their original bottle.
ext
®
n
ontour
ext
• Use only approved equipment from the manufacturer or certified
body such as UL or TUV.
• It is advisable to avoid use of electronic devices in very dry
environments, especially if synthetic materials are present.
• Your C
®
n
meter has been preset and locked to
ontour
ext
display results in mmol/L (millimoles of glucose per liter
of blood).
Results in mmol/L will always have a decimal point;
❖
Results in mg/dL will never have a decimal point.
❖
Example:
Check your display screen to be sure the results are shown
❖
the right way. If not, contact Customer Service. See the back
cover for contact information.
• The C
n
blood glucose monitoring system has a
®
ontour
ext
measuring range of 0.6 mmol/L to 33.3 mmol/L.
For results over 33.3 mmol/L or under 0.6 mmol/L:
❖
If your blood sugar reading is under 0.6 mmol/L, the "Result
о
Under 0.6 mmol/L" screen will instruct you to "Follow Medical
Advice Immediately." Contact your health care professional.
If your blood sugar reading is over 33.3 mmol/L the "Result
о
Over 33.3 mmol/L" screen will instruct you to "Wash hands or
test site. Repeat test with new strip". If results are still over
33.3 mmol/L, "Follow Medical Advice Immediately"
is displayed.
or
iv
9/16/16 3:26 PM