Step 4 Pull the wire gently to confirm that the wire may not be disconnected.
Step 5 Insert the easy displacement connector into sensor connector No. 1 of this device.
Maximum 4 safety liquid leak sensor can be connected.
Step 6 Depending on No. of the safety liquid leak sensor connecting, set the sensor connection
number setting switches as table below.
Sensor connection
number setting
switches
© Panasonic Industriar Devices SUNX Co., Ltd. 2016
easy insulation displacement connector
A
be reused.
easy insulation displacement connector
Use optional
pcs. / set) and CN-EP3 (for PFA cable, 5 pcs. / set) or recommended parts.
<Recommended easy insulation displacement connector>
e-CON: [By Tyco Electronics Japan G.K.]
• For PVC cable: 1473562-4
• For PFA cable: 3-1473562-4
● When connecting to the sensor connector, be sure connect the safety liquid
leak sensors from No. 1 in order. If you connect in disorder, the controller be-
comes lockout status and does not work.
● In case connecting PNP output type of safety liquid leak sensor, set to PNP side
the control output polarity selection switch of this device. Also, in case connect-
ing NPN output type set to PNP side the control output polarity selection switch
of this device.
Switch No.
5
6
ON
ON
ON
OFF
OFF
ON
OFF
OFF
Wire
Pull it gently
Sensor connector
Insert
Wired easy insulation displacement connector
on the safety liquid leak sensor.
No. of the sen-
sor connecting
7
8
ON
ON
1 unit
ON
OFF
2 units
OFF
ON
3 units
OFF
OFF
4 units
that has been crimped once cannot
CN-EP2 (for PVC cable, 5
Wiring
19