The sample container is filling too
quickly.
23
IOM-195
MODEL: INLOOP™ ACE
Table 2: inLoop™ ACE Troubleshooting (Continued)
Possible Causes
The inLoop™ ACE is not collecting the
correct sample volume.
The inLoop™ ACE may be set at a faster
sampling frequency than desired.
The relief on the inLoop™ ACE may be
set too low.
REV: B
13839 West Bellfort Street, Sugar Land, TX 77498
Solutions
Welker recommends the Welker
Checkpoint™ Sample Bite Verification
Panel for verifying the sample volume of
the inLoop™ ACE. See Section 2.2, Setting
the Sample Volume, for instructions on
adjusting the sample volume.
Adjust the inLoop™ ACE to sample at the
desired rate. Ensure that the calculations
used to determine the sample frequency
are correct (Figure 8).
Check the setting on the relief and adjust
as necessary.