Illumina DRAGEN Server v3 Site Prep and Installation Guide
Power Cords
The DRAGEN Server v3 has international standard IEC 60320 C13 receptacles and is shipped with one
region/country-specific power cords.
Hazardous voltages are removed from the DRAGEN Server v3 only when the power cords are disconnected
from the AC power source.
To obtain equivalent receptacles or power cords that comply with local standards, consult a third-party
supplier such as Interpower Corporation (www.interpower.com).
CAUTION
Never use an extension cord to connect the DRAGEN Server v3 to a power supply.
Fuses
The DRAGEN Server v3 contains no user-replaceable fuses.

Environmental Considerations

The following table specifies temperature, humidity, and other environmental considerations for safely housing
the instrument.
Element
Specification
Temperature
Maintain a server room temperature of 0°C to 40°C. This temperature is the operating
temperature of the DRAGEN Server v3. Do not allow the ambient temperature to vary by
more than ±2°C.
Humidity
Maintain a noncondensing relative humidity between 10–95%.
Altitude
Locate the DRAGEN Server v3 at an altitude below 2,000 meters (6,500 ft).
Air Quality
Operate the server in a data center environment with air filtration per ISO 14644-1 Class 8
with a 95% upper confidence limit.
Ventilation
Consult your facilities department for ventilation requirements sufficient for the level of heat
output expected from the DRAGEN Server v3.
Location
Operate the server in an indoor environment only. The server is not suitable for use in
locations where children are likely to be present.
Heat Output
Measured Power
1100 Watts
Noise Output
A measurement of ≤ 70 dB is within the level of a normal conversation at a distance of approximately 1 meter
(3.3 feet).
Acoustical Performance
Idle at 25°C Ambient
Max. Loading at 35°C Ambient
Document # 1000000097923 v02
For Research Use Only. Not for use in diagnostic procedures.
Thermal Output
4100 BTU/hr
Noise Output (dB)
36
<75
3