Status coding for ULTRAMAT/OXYMAT/CALOMAT:
Dec Hex
11
0B
Component missing
12
0C
Device error
31
1F
Out of order (equivalent value,
last valid value, through connect
value)
Dec Hex
64
40
Function control
71
47
Measured value is bad,
last valid value
75
4B
Measured value is bad,
equivalent value
79
4F
After switching on the power,
initial value
80
50
Measured value is out of range
Dec Hex
128
80
o.k., normal operation
132
84
Active block alarm, parameter
has been changed
137
89
Lower warning limit exceeded
138
8A
Upper warning limit exceeded
141
8D
Lower alarm limit exceeded
142
8E
Upper alarm limit exceeded
164
A4
Maintenance request
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
Status coding for "bad quality"
Cause
Change configuration, install analyser including this
component
Check with SIMATIC PDM / change analyzer
Depending on the paramterization. Switch in mode AUTO
Status coding for "uncertain quality"
Cause
Warm up, AUTOCAL, in situ service
Check with SIMATIC PDM / change analyzer
Check with SIMATIC PDM / change analyzer
Reject the measured value
Check with SIMATIC PDM / ambient temperature too high
Status coding for "good quality"
Cause
You can evaluate the measured values
Disappears after 10 seconds
user programm
user programm
user programm
user programm
Check with SIMATIC PDM
Remedy
Remedy
Remedy
95