Siemens ULTRAMAT 6 Manual page 114

Hide thumbs Also See for ULTRAMAT 6:
Table of Contents

Advertisement

Available languages

Available languages

5.2. Setting the Station Address ULTRAMAT 23
Note
The basic principles of operation are explained in chapter 5 of the ULTRAMAT 23
device manual.
The station address can be set in the main menu under Configuration with the Special Functions submenu
item. The level 2 code is required for this.
Codes/Language
AUTOCAL-drift
ELAN / PROFIBUS
Fact.Data/Res/Units
ELAN Parameter
ELAN ext.Interfer.
PROFIBUS Parameters
PROFIBUS
Address
7
Warning
Changing the station address can affect communication on the bus.
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
: 126
Select the ELAN/PROFIBUS subitem
from this menu
Next select
PROFIBUS Parameters subitem
The PROFIBUS station address can be
set in this dialog.
Addresses between 0...126 can be set.
114