Installation Of Sensor Cartridge; Registration Of The Sensor Cartridge; Calibration - Intec PolyGard 2 SC2 User Manual

Toxic (e-), combustible (p-) and infrared (i-co2) gases
Table of Contents

Advertisement

SC2-AT6 – UserManual
Specifications subject to change without notice. | Tox_D_1119, Ex_D_1119, CO2_D_0320 | USA 200413 | Page 7 of 17

4.1 Installation of Sensor Cartridge

The Sensor Cartridges are provided already installed to the PG2 devices and capped to protect against dirt and
damage. Replacement sensors are available individually:
Check gas type, range and calibration date of Sensor Cartridge.
Define installation place on the housing of the basic or remote sensor and break out knockouts.
Tighten the Sensor Cartridge with M32 hexagon lock nut.
Plug in the Sensor Cartridge at X2 or X3 of the sensor board. Observe plug polarity, the plug must engage.

4.2 Registration of the Sensor Cartridge

Registration and addressing of the field bus can be read in the manual for the specific PG2 transmitter or controller.

5 Calibration

The service tool STL6 or DPT6 (DGC6 EasyConf software) are available for convenient on-site calibration. See their
respective datasheet.
A sensor exchange program is also available, consult your account executive.
Prior to calibration the Sensor Cartridge must be supplied with power voltage without interruption for
warm-up and stabilization. The warm-up time depends on the sensor element and is shown in the
following tables:
Gas type
Ammonia
Chlorine
Hydrogen cyanide
Formaldehyde
Carbon monoxide
Ozone
Sulfur dioxide
Hydrogen sulphide
Nitrogen dioxide
Oxygen (-A/-A2)
Gas type
All (combustible) pellistor
sensors
Gas type
Carbon dioxide
Please observe proper handling procedures for compressed gas and test gas bottles.
Test gas can be toxic; never inhale it!
• Symptoms: Dizziness, headache and nausea.
• Procedure if exposed: Remove victim to fresh air at once, seek medical attention.
INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com
Formula Calibration with
replacement gas
(Sensitivity in %)
NH
-
3
Cl
-
2
HCN
-
CH
O
-
2
CO
-
O
-
3
SO
-
2
H
S
-
2
NO
-
2
O
-
2
Stabilization time
to spec (min)
60
Stabilization time
to spec (min)
CO
10
2
Warm-up
Flow rate
Calibration interval
(h)
(ml/min)
in months
18
300
6
150
6
300
24
300
See Specification
1
150
section later in this
6
150
document
6
300
6
150
6
500
1
150
Warm-up
Flow rate
Calibration interval
(h)
(ml/min)
in months
24
150
See Specification
section later in this
document
Warm-up
Flow rate
Calibration interval
(h)
(ml/min)
in months
24
1500
See Specification
section later in this
document