Setting The Dosage In Continuous Operation - Bossard KDG 1000 Series Operating Instructions Manual

Dosing unit
Table of Contents

Advertisement

Dosing
1
3
5
6

Setting the dosage in continuous operation

(Continuous mode)
• Install device (see page 13)
• Set the device to the default settings (see page 13)
• Fill cartridge (see page 17)
• Switch on the power switch (1)
1.
Netzschalter O n/O ff
• Set operating mode to «Continuous»
2.
Druckregler
3.
Vakuumregler
4.
LE D
• Remove the valve gate (26)
5.
Display
6.
Kartuschenanschluss
• Press and hold the foot switch (18)
7.
Menutaste
8.
E ntertaste
9.
Pfeiltaste auf
• Use the pressure regulator (2) to increase from 0
10.
Pfeiltaste ab
11.
Netzanschluss
to the required quantity
12.
Schnittstelle ext. Steuerung
13.
Initiatoranschluss
14.
Druckluftanschluss
• For low-viscosity products, set the vacuum to prevent the product from escaping
15.
Leistungsschild
from the cartridge when removing the valve gate (26) and mounting the dosing needle (27).
Allow 1 or 2 drops to flow out of the dosing needle.
Then use the vacuum regulator (3) to slowly adjust the vacuum.
Turn counterclockwise:
Turn clockwise:
Depending on the viscosity and size of the cartridge, the vacuum will have to
be adjusted again
If the vacuum is set too high – when dosing with no piston – the medium can escape from the
cartridge into and damage the dosing unit.
Press the foot switch (18) repeatedly with the pressure regulator (2) to increase the dosing pressure gradu-
ally from 0 until the required quantity is reached.
If the required dosage is not reached, repeat the repeat the steps using larger or smaller dosing nozzles.
4
7
9
2
8
10
The back pressure increases.
The back pressure decreases.
15