Alarm Conditions
17
Normal Alarms
When a paired monitor sends out an alarm call, the Connect-iT's "Monitor" LED will turn solid red. The "Alarm"
LED will flash every second. The wired output will activate every 2 minutes until the monitors alarm condition has
been cleared. See page 16 for resetting an alarm.
Common Fault Conditions
Radio Fail Alarm:
If the Connect-iT has been paired to a device with "Safelink" enabled (Type code over 128) it will expect to see the
heartbeat signal from the monitor every 8 seconds. If it does not see any heartbeat signals for 2 minutes it will raise
a Radio Failure alarm. In this condition the wired output will activate, the "Monitor" LED will blink every 2 seconds
and the "Alarm" LED will blink every 2 seconds.
Monitor Battery Low / Very Low:
When a paired monitors internal battery gets 'low', the "Monitor" led will begin to blink red every second.
When a paired monitors internal battery gets 'very low', the "Monitor" LED will blink every second, the "Alarm"
LED will start to blink every 2 seconds and the wired output will activate.
Connect-iT Backup Battery Low / Very Low:
When running on the battery backup, the "Power" LED will turn from solid green to blinking green every 4 seconds.
When the backup battery begins to get low the "Power" LED will start to blink every second.
When the backup battery gets 'very low' the wired output will activate, the "Monitor" LED will begin to blink every
second and the "Alarm" LED will flash.
Need help?
Do you have a question about the Companion Mini and is the answer not in the manual?
Questions and answers