Chapter 2 Methods
2

Set up a run

Set up a run
12
1. Switch the power switch located at the rear of the unit to the On position.
2. Touch Set up run.
3. In the Method Type screen touch either Constant voltage, Constant current,
or Constant power.
Figure 5 Method Type screens
4. Edit the run parameters (Volts, Time, Amps, and Watts) by touching the
parameter on the screen to bring up the Run parameter edit screen.
Default maximum values are 300 V, 500 mA, and 120 W.
Note: When using constant voltage or constant power, it is recommended
that the maximum default settings not be changed to avoid unwanted power
limitations during the run.
5. Enter the parameter value required, then touch Enter to advance with the newly
changed run parameter displayed.
6. (Optional) Add or delete extra steps by touching Actions.
PowerEase
Touch 120W Power Supply User Guide