Page 1
Electrophoresis Power Supply EC300XL Operating and Maintenance Manual 7217581 Rev. 0 Visit us online to register your warranty www.thermoscientific.com/warranty...
Page 2
Preface MANUAL NUMBER 72175881 5/3/12 Transfer to Marietta (was LTM1596X2 4/7/10) ECR/ECN DATE DESCRIPTION Thermo Scientific EC300XL...
Page 4
Always use the proper protective equipment (clothing, gloves, goggles, etc.) 4 Always dissipate extreme cold or heat and wear protective clothing. 4 Always follow good hygiene practices. 4 Each individual is responsible for his or her own safety. Thermo Scientific EC300XL...
Page 5
We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service...
Introduction Section 1 Thank you for selecting a Thermo Scientific EC300 Electrophoresis power supply. This manual describes the operation of the EC300. The power supply that you have purchased is the most productive and easy-to-use unit available anywhere. This manual should answer any questions that might arise in operating your power supply;...
This applies only to the EC300 when used as specified in the documenta- tion, in its intended applications, with Thermo Scientific approved elec- trophoresis apparatus only. Usage in any other manner may not provide suitable protection.
3-prong grounded AC outlet, using the AC cord provided with the unit only. Black negative 3 digit LED display jacks Up/Down arrows Select key (Set) Constant key (K) Run key Red positive Blue On/Off key jacks Figure 2-1. Control Panel Thermo Scientific EC300XL...
The limit parameter cannot be changed during a run. The time parameter displays elapsed time for an untimed run, and time left for a timed run. EC300XL Thermo Scientific...
75W. When starting a run, the power supply calculates a maximum limit parameter, above which the power supply maximum power output rating would be exceeded. If this value is lower than the limit setpoint entered, the EC300 will use this lower limit parameter. Thermo Scientific EC300XL...
AC power, the run was not completed since PF was not enabled. Loss of AC power; run not timed. The EC300 detected loss of AC power. 06-99 The EC300 detected an internal error condition. Contact Technical Services for instructions. Thermo Scientific EC300XL...