WAITING FOR PARAFFIN
Po 1• •
C XX• • •
TR=00:00
Thermo Scientific Spin Tissue Processor Microm STP 120
CAUTION
Before executing the first real process with reagents and paraffin,
remember that paraffin can sometimes take a long time to melt.
After a long period of disconnection (for example the first start or
after power failure), the basket will not descend into the paraffin
bath (even though this is programmed) until the default safety
period has elapsed.
When testing the programming, skip paraffin bath positions 11 and
12 (and 10 if working with two baskets) by assigning them a time
period of 00:00.
If you program a time period of more than 00:00 in positions 11
and 12 (10, if working with three paraffin baths), the screen will
display this message, where XX is the number of the reagent
vessel programmed prior to paraffin bath 11 (10 if a third bath is
installed).