Electromagnetic Compatibility Information
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such
an environment.
IMMUNITY test
IEC 60601
test level
± 8 kV
Electrostatic
contact
discharge (ESD)
±2 kV,±4 kV,
±8 kV,
IEC 61000-4-2
±15 kV air
Electrostatic
± 2 kV ,
transient/burst
100kHz, for
IEC 61000-4-4
AC power port
Surge
±0.5kV, ±1kV
IEC 61000-4-5
(differential
mode)
Voltage dips,
0 % UT;
short interruptions
0,5 cycle
and voltage
At 0°, 45°, 90°,
variations on
135°, 180°, 225°,
power supply
270° and 315°
input lines
0 % UT; 1 cycle
IEC 61000-4-11
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
Power frequency
30 A/m; 50Hz
(50/60 Hz)
or 60Hz
magnetic field
IEC 61000-4-8
EssentiA
Compliance
Electromagnetic environment
level
-guidance
± 8 kV
Floors should be wood, concrete
contact
or ceramic tile. If floors are
±2 kV,±4 kV,
covered with synthetic material,
±8 kV,
the relative humidity should be
±15 kV air
at least 30%.
± 2 kV ,
Mains power quality should be that
of a typical commercial or hospital
100kHz, for
environment.
AC power port
±0.5kV, ±1kV
Mains power quality should be that
.
of a typical commercial or hospital
(differential
environment.
mode)
0 % UT;
Mains power quality should be that
0,5 cycle
of a typical commercial or hospital
At 0°, 45°, 90°,
environment.
135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
30 A/m; 50Hz
Power frequency magnetic fields
or 60Hz
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.