I - INTRODUCTION
Congratulations on the purchase of your new P
S
™/P
S
L
OLO
IEZOTOME
OLO
To maximise use of this cutting-edge product for
as long as possible, please read these instructions
carefully before switching on, using or servicing the
device.
Sentences containing the symbol
particular importance.
Designed for clinical use:
Created by SATELEC, P
a device designed to perform dental surgery including
osteotomies, osteoplasties, periodontal and implant
surgery...
II - WARNINGS
CAUTION:
United States Federal Law restricts the use of this
device solely to qualiÀ ed, trained and competent
dental health practitioners or under their
supervision.
The surgical tips designed for P
S
L
are not compatible with the P
OLO
ED
the I
C
, and vice versa.
MPLANT
ENTER
P
S
/P
IEZOTOME
OLO
IEZOTOME
adapted to P
2 or I
IEZOTOME
To reduce the risk of accidents, the precautions
stipulated below must be taken:
Device operators:
- Use of the device is restricted solely to graduate
dental health professionals, who are competent and
qualiÀ ed within the usual context of their activities.
- If you have received this device by mistake, please
contact the device supplier to arrange for its return.
device.
ED
are points of
S
/P
IEZOTOME
OLO
IEZOTOME
S
IEZOTOME
OLO
IEZOTOME
S
L
can only take tips
OLO
ED
C
2.
MPLANT
ENTER
Interactions:
Interference may occur when used on
patients with cardiac pacemakers.
This system emits electromagnetic À elds,
IEZOTOME
which means there are some potential risks. The
malfunctioning of implantable devices such as cardiac
pacemakers and ICDs (implantable cardioverter
deÀ brillator) is possible:
- Ask patients and users if they have an implanted
device before using this product. Explain the
circumstances to them.
- Weigh the risks and beneÀ ts and contact your
patient's cardiologist or appropriate qualiÀ ed
healthcare professional prior to performing the
S
L
is
treatment.
OLO
ED
- Keep this product away from implanted devices.
- Make appropriate emergency provisions and take
immediate action if patients become ill.
- Symptom including a raised heartbeat, irregular
pulse, and dizziness may signal problems with a
cardiac pacemaker or ICD.
Electrical connection:
- Ask an approved dental installation technician to
connect your device to the electrical network.
- Warning: To prevent any risk of electric shock, this
device shall only be connected to a supply network
À tted with a protective ground.
/P
- The electrical network to which the device is
IEZOTOME
and
connected shall comply with the standards in force
in your country.
- If loss of electrical power during use is likely to
generate an unacceptable risk, the device must be
connected to a suitable power source (e.g. UPS).
- Do not attempt to connect to the device any other
accessories than those supplied by SATELEC.
Overheating, burns:
To avoid the risk of burns or of overheating:
- Please refer to the clinical booklet provided with
the device.
- T he irrigation Á ow must be used all the time
during the clinical act (please refer to the clinical
booklet).
- Please replace the tips when the cutting part
becomes blunt or when the diamond coated part
5