shut down the device with the mains isolating switch
(Fig. 2-5) then switch the device back on.
When the ultrasound function is activated and
the handpiece is not connected or is not properly
connected, symbol 2 (Chap XVII - Symbols) lights up
(Fig. 9-1) and 4 beeps sound. We recommend that you
connect the handpiece or check that it is correctly
connected to the device.
IX - USING THE DEVICE
Important:
- Do not disconnect the handpiece cable when the
device is on and the pedal is pressed down.
- Do not tighten or loosen the tips when the
handpiece is activated.
- All accessories used shall be previously cleaned,
disinfected and sterilized.
- For your safety and the safety of your patient,
the device shall only be used with accessories
supplied or recommended by SATELEC.
- Before and after each use, check the device and its
accessories to make sure there are no problems.
- If necessary, do not use the device and replace
any faulty parts.
Operate the device as follows:
- Make sure that the device is correctly plugged in and
corrected insulated.
- Set the mains isolating switch (Fig.2-5) to position
I (ON).
- Hang the bottle or pouch containing physiological
saline solution or sterile water from the bracket.
- Lift the lid of the irrigating pump.
- Horizontally insert an irrigation line cassette into
the purpose-designed housing.
- Lower the lid.
- Connect the irrigation line perforator to the bottle
or pouch containing physiological saline solution or
sterile water.
- Set the required parameters (irrigation Á ow rate,
programme), see chapter VII.
Note:
Automatic storage allows you to save the display
parameters used previously when switching between
the different programmes.
This sequence is carried under controlled aseptic
conditions for the safety of your patients. You may
require the help of an assistant.
- Connect the handpiece cable to the device connector
(Fig. 1-7).
- Screw the selected tip onto the handpiece using a
wrench (according to the recommendations of the
clinical manual).
- Attach the irrigation line to the cable using line
clips.
- Attach the end of the irrigation line to the handpiece.
- Prime the irrigation system by pressing the Á ush
button on the screen (Fig. 3-3)
- When the irrigation solution arrives at the tip of the
tip, release the pedal.
- Use the device in compliance with good dentistry
practice.
X - SWITCHING OFF THE
DEVICE
At the end of the dental procedure:
- Remove the bottle or pouch containing physiological
saline solution or sterile water from the bracket.
- Remove the irrigation line perforator from the bottle
or pouch.
- Immerse the irrigation line perforator in a container
of distilled water.
- Rinse the irrigation line and the handpiece by
activating the Á ush function until the receptacle
and the irrigation line are completely empty.
- Set the device to OFF (O).
- Disconnect the handpiece from the device.
- Remove the irrigation line clips.
- Disconnect the irrigation line from the handpiece
and throw it in a bin used for the disposal of used
medical items (irrigation line can only be used
once). See chapter 11.4.