Emc Table - i-SENS VetMate User Manual

Blood glucose monitoring system
Hide thumbs Also See for VetMate:
Table of Contents

Advertisement

Available languages

Available languages

EMC Table

The following tables contain the Manufacturer's declaration and
additional information required by IEC 60601-1-2:2014 (Fourth
Edition).
Basic EMC standard
Phenomenon
or test method
CISPR 11:2015
Mains terminal
+A1:2016
disturbance voltage
EN 55011:2016
+A1:2017
CISPR 11:2015
Radiated
+A1:2016
disturbance
EN 55011:2016
+A1:2017
Harmonic Current
IEC 61000-3-2:2014
Emission
EN 61000-3-2:2014
Voltage change,
IEC 61000-3-3:2013
Voltage fluctuations
EN 61000-3-3:2013
and Flicker Emission
Electrostatic
IEC 61000-4-2:2008
Discharge Immunity
EN 61000-4-2:2009
IEC 61000-4-3:2006
Radiated RF
+A2:2010
Electromagnetic
EN 61000-4-3:2006
Field Immunity
+A2:2010
Immunity to
IEC 61000-4-3:2006
Proximity Fields
+A2:2010
from RF wireless
EN 61000-4-3:2006
Communications
+A2:2010
Equipment
46 www.i-sens.com
Port
Test level/requirement
tested
AC Main
Group1, Class B
Enclosure
Group1, Class B
AC Main
Class A
Pst: 1
Plt: 0.65
AC Main
Tmax: 0.5
dmax: 4 %
dc: 3.3 %
± 8 kV/Contact
Enclosure
± 2, ± 4, ± 8, ± 15 kV/Air
10 V/m,
Enclosure
80 MHz - 2.7 GHz
80% AM at 1 kHz
Table 9 in
Enclosure
IEC 60601-1-2: 2014
Basic EMC standard
Phenomenon
or test method
Electrical Fast
IEC 61000-4-4:2012
Transient/Burst
EN 61000-4-4:2012
Immunity
IEC 61000-4-5:2014
Surge Immunity
EN 61000-4-5:2014
Immunity to
Conducted
IEC 61000-4-6:2013
Disturbances
EN 61000-4-6:2014
Induced by RF fields
Power Frequency
IEC 61000-4-8:2009
Magnetic Field
EN 61000-4-8:2010
Immunity
IEC 61000-4-11:
2004
Voltage dips
EN 61000-4-11:
2004
Voltage
IEC 61000-4-11: 2004
interruptions
EN 61000-4-11: 2004
Port
Test level/requirement
tested
± 2 kV, 100 kHz
AC Main
repetition frequency
Line to Line
± 0.5 kV, ± 1 kV
AC Main
Line to Ground
± 0.5 kV, ± 1 kV,
± 2 kV
3 V
0.15 MHz - 80 MHz
6 V in ISM and
AC Main
amateur radio bands
Between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
30 A/m
Enclosure
50 Hz and 60 Hz
0 % U
: 0.5 cycle
T
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
AC Main
0 % U
; 1 cycle and
T
70 % U
; 25/30 cycles
T
Single phase: at 0º
AC Main
0 % U
; 250/ 300 cycle
T
www.i-sens.com 47

Hide quick links:

Advertisement

Chapters

Table of Contents
loading