Comparing the Control Solution Test Results

The test result of each control solution should be within the range
printed on the label of the test strip vial. Repeat the control solution
test if the test result falls outside of the range. Out of range results
may occur in the following situations:
Situations
• When the control solution bottle was not
shaken well,
• When the meter, test strip, or the control
solution were exposed to high or low
temperatures,
• When the first drop of the control solution
was not discarded or the tip of the bottle was
not wiped clean,
• When the control solution is past the
expiration date printed on the bottle,
• When the control solution is past its discard
date (the date the bottle was opened plus
three (3) months),
• When the control solution is contaminated.
If results continue to fall outside the range printed on the test
strip vial, the VetMate Test Strip and Meter may not be working
properly. Do not use your system and contact Customer Service
at 1-800-429-5001 Mon–Sat, 9 am–9 pm EST. At all other times,
contact your veterinarian/veterinary surgeon.
Note:
The VetMate Control Solution can be purchased
separately. Please contact Customer Service at 1-800-429-
5001 Mon–Sat, 9 am–9 pm EST. At all other times, contact your
veterinarian/veterinary surgeon.
24 www.i-sens.com

Testing

Using the Lancing Device

You will need a lancing device in order to collect a blood
sample.
You may use the lancing device included in the VetMate
Blood Glucose Monitoring System or any other medically
approved lancing device.
Do This
Repeat the
control solution
test by referring
to the 'Notes'
on page 20.
• Use a soft cloth or tissue to wipe the lancing device.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
Adjustable Tip
Lancet Holder
Lancet
If necessary, a small amount of alcohol on a soft cloth or
tissue may be used.
Caution:
To avoid infection when drawing a sample, do not
use a lancet more than once, and:
• Do not use a lancet that has been used by other dogs/cats.
• Always use a new sterile lancet.
• Keep the lancing device clean.
Ejection System
Release Button
Sliding Barrel
www.i-sens.com 25