i-SENS VetMate User Manual page 13

Blood glucose monitoring system
Hide thumbs Also See for VetMate:
Table of Contents

Advertisement

Available languages

Available languages

Step 2
Shake the VetMate Control
Solution bottle before each test.
Remove the cap and squeeze the
bottle to discard the first drop.
Then wipe the tip with a clean
tissue or cloth. After the
symbol
appears on the display, apply
the solution to the narrow edge
of the test strip until the meter
beeps. Make sure the confirmation
window fills completely.
Note:
The meter may switch off if the control solution sample
is not applied within 2 minutes of the
on the screen. If the meter turns off, remove the strip, reinsert,
and start from step 1.
Step 3
A test result will appear after the meter
counts down from 5 to 1.
When 'check' is displayed, the result is
stored in the meter's memory but it is not
included in the averages.
22 www.i-sens.com
symbol appearing
Step 4
Compare the result displayed on the
meter to the range printed on the test
strip vial.
The result should fall within the range.
Used strips should be discarded safely
in disposable containers.
Caution:
• If the results are not within the range on the test strip vial, the
meter and test strip may not work properly. Then, stop using the
meter and contact Customer Service: 1-800-429-5001 Mon–Sat,
9 am–9 pm EST.
• The range printed on the test strip vial is for the VetMate Control
Solution only. It does not have any connection to the blood
glucose level of your dog/cat.