i-SENS VetMate User Manual page 24

Blood glucose monitoring system
Hide thumbs Also See for VetMate:
Table of Contents

Advertisement

Available languages

Available languages

For cats
Slope
Y-Intercept
Correlation coefficient (r)
Number of measurements
Range tested
Accuracy results for glucose concentration < 100 mg/dL
Within ± 5 mg/dL
Within ± 10 mg/dL
114/210 (54.3 %)
189/210 (90 %)
Accuracy results for glucose concentration ≥ 100 mg/dL
Within ± 5 %
Within ± 10 %
204/390 (52.3 %)
331/390 (84.9 %)
Accuracy results for glucose concentration between 28.9 mg/dL
and 496.3 mg/dL
Within 15 mg/dL or 15 %
595/600 (99.2%)
44 www.i-sens.com
1.0533
4.9766
0.9935
600
28.9–496.3 mg/dL
Within ± 15 mg/dL
209/210 (99.5 %)
Within ± 15 %
386/390 (99 %)
PRECISION: Precision studies were performed in a laboratory
using the VetMate BGM Systems.
For dogs
Within Run Precision
Interval 1. 30–50
36.9 mg/dL
Interval 2. 51–110
60.1 mg/dL
Interval 3. 111–150
123.7 mg/dL SD = 3.3 mg/dL CV = 2.6 %
Interval 4. 151–250
187.6 mg/dL SD = 8.5 mg/dL CV = 4.5 %
Interval 5. 251–400
286.9 mg/dL SD = 11.2 mg/dL CV = 3.9 %
For cats
Within Run Precision
Interval 1. 30–50
34.6 mg/dL
Interval 2. 51–110
64.9 mg/dL
Interval 3. 111–150
134.2 mg/dL SD = 4.4 mg/dL CV = 3.3 %
Interval 4. 151–250
191.1 mg/dL SD = 8.6 mg/dL CV = 4.5 %
Interval 5. 251–400
304.6 mg/dL SD = 12.2 mg/dL CV = 4.0 %
For dogs and cats
Between Run Precision
Interval 1. 30–50
42.2 mg/dL
Interval 2. 96–144
121.5 mg/dL SD = 4.7 mg/dL CV = 3.9 %
Interval 3. 280–420
357.8 mg/dL SD = 9.8 mg/dL CV = 2.7 %
SD = 2.3 mg/dL
SD = 4.2 mg/dL
SD = 2.7 mg/dL
SD = 2.6 mg/dL
SD = 2.3 mg/dL
www.i-sens.com 45

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

Table of Contents