Fetco CBS-2111XTS Users Manual And Operator Instructions page 11

Extractor touch screen models
Table of Contents

Advertisement

<
OUTPUTS
<.
Brew Valve
<
Calibrate/Test
«
<
OUTPUTS
<
Brew Valve
<.
Calibrate/Test .>
Cal
Test
<<
<
OUTPUTS
<
Brew Valve
<
Calibrate/Test
.
Cal
Test
.
Press to Cal/Test
<<
BASKET: IN PLACE
TEMP: READY
<
OUTPUTS
<
Brew Valve
<
Calibrate/Test
.
Cal
.
To Stop press Cal
Cal
Time
Rem.
0:60
CAL IN PROGRESS
<
OUTPUTS
<
Brew Valve
<
Flow Rate
.
1218 mlpm
.
|________[]_____|
-
..
500
2000...
Go to fetco.com for the latest versions of all information
EXAMPLE: Calibrate BREW VALVE.
1) Enter PROGRAMMING menu
(See page 5: INSTRUCTIONS TO ACCESS PROGRAMMING MENUS)
>
Scroll to "OUTPUTS" in PROGRAMMING
.>
>
Scroll down to BREW VALVE
»
(
The following areas may be accessed in this PROGRAMMING position:
BREW VALVE, BYPASS VALVE, Fill Valve, Heater, Buzzer, Screen. To access
these: with the 2
right/left arrows. Only the Brew and Bypass valves can be calibrated
Scroll down to "Calibrate/Test"
>
(
The following areas may be accessed in this PROGRAMMING position:
>
Calibrate/Test and Flow Rate. To access these: with the 3
shown—scroll through the positions with the right/left arrows. CBS-2111
automatically calibrates based on the flow entered. The highlighted FLOW RATE is
used to enter the flow rate and for Service and to fine tune the flow
»
Scroll down to "Cal
CAL
>
>
>
Have 3 Liter/1Gallon measuring container under brew basket
.
.
TEST
In "Cal
PRESS CAL and remove finger—do not hold
CAL
>
>
The CBS-2111 will dispense for exactly one minute and automatically stop.
>
.
.
Carefully measure the amount of water, in milliliters, that was dispensed
This is the calibration test flow
The calibration results are enetered into the slider that automacially displays
ENTER THE CALIBRATION TEST FLOW RESULTS
The calibration is adjusted by the CBS-2111XTS software by entering the
amount of water into the slilider that was dispensed in 60 seconds.
>
The slider shows 1218 mlpm (milliters per minute)-enter your results.
>
Using the + and – icons, enter the amount of water dispensed on the slider.
>
Enter in milliliters (multiply fluid ounces by 29.6 to obtain milliliters)
.
..
Scroll UP and OUT and save calibration as directed by screen prompts.
+
Brew valve is automatically calibrated on saving the test flow entered.
EXIT TO COMPLETE CALIBRATION
Page 11
nd
level highlighted as shown—scroll through the positions with the
Test"
Test
User Guide & Operator instructions
)
rd
level highlighted as
.)
P171 NEW MARCH 2018