Electrosurgical unit

Advertisement

INSTRUCTIONS
ELECTROSURGICAL UNIT
WARNING
The user of this equipment should be thoroughly trained in the applicable procedure.
Furthermore, failure to read and thoroughly understand the contents of this instruction
manual may result in serious injury to the patient and/or user. It is essential to follow
the instructions contained in this and other manuals which pertain to any equipment
and accessories used in conjunction with this equipment. Possible injuries related
to endoscopic procedures may include electric shock, explosion, burns, perforation,
hemorrhage, etc.
Failure to follow these instructions may also result in damage to and/or malfunction
of this equipment.
CAUTION
Federal (USA) law restricts this device to sale by or on the order of a physician.
Non-Controlled Document

Advertisement

Table of Contents
loading

Summary of Contents for Olympus PSD-10

  • Page 1 INSTRUCTIONS ELECTROSURGICAL UNIT WARNING The user of this equipment should be thoroughly trained in the applicable procedure. Furthermore, failure to read and thoroughly understand the contents of this instruction manual may result in serious injury to the patient and/or user. It is essential to follow the instructions contained in this and other manuals which pertain to any equipment and accessories used in conjunction with this equipment.
  • Page 2 If you have any questions concerning the material contained in this manual or concerning the operation or safety of the equipment, please contact your Olympus representative or the nearest Olympus Service Center. Non-Controlled Document...
  • Page 3: Table Of Contents

    Main Specifications 1 - 3 Safety Precautions DESCRIPTION OF CONTROLS CONSTRUCTION STANDARD SET PREPARATION FOR USE Installation of PSD-10 5 - 2 Connection to AC Power 5 - 3 Connection of Footswitch 5 - 4 Connection of S-P Cord INSPECTION OF ELECTROSURGICAL SYSTEM...
  • Page 4: Features And Main Speclflcatlons

    Features 1. Simplified pre-procedural test, using a Check Lead, of the power output of the PSD-10 alone as well as that of the entire electrosurgical system includ- ing the endoscope and endoscopic electrodes. Test results confirmed by TEST “OK” or “FAULT” indicator lights on the front panel.
  • Page 5: Main Specifications

    1-2 Main Specifications Endoscopic electrosurgery Applicable field Applicability Olympus fiberscopes suitable for electrosurgery Compatible endoscopes Waveform BLEND COAG Output (watts) Output Radio-frequency Max. rated output Current (Load: 5OO CUT: 80 W BLEND: 60W COAG: 40 W Output setting Max. open-curcuit output voltage: less than 1800 Vp-p...
  • Page 6 Classification Degree of protection Type BF (Body Floating) equipment (Electro- against electric shock medical equipment) Degree of protection The PSD-10 should never be used where there is a risk against explosion of flammable gases. Defibrillator-protected equipment Other Non-Controlled Document...
  • Page 7 10 – 40°C (50 – 104ºF) 30 – 75% Relative humidity Atmospheric pressure 700 –1060hPa (700 – 1060 mbar) Dimensions 290 mm (W) x 162 mm (H) x 300 mm (D) max., & Weight 8 kg or 17.6 lb (PSD-10 main body) Non-Controlled Document...
  • Page 8: Safety Precautions

    1 - 3 S a f e t y P r e c a u t i o n s The PSD-10 should be used only in a medical facility under the supervision of a trained physician. The PSD-10 is designed (floating) to function integrally with insulated patient attachments (active electrode and patient plate) to prevent operator/patient shock.
  • Page 9 Do not contact these cords with the patient or metallic portion of an operating table, etc. The PSD-10 should not be used in conjunction with electrical apparatus whose safety against leakage current is not guaranteed. Radio-frequency and spark discharge “noise” may adversely affect other electro-medical equipment being used in the proximity.
  • Page 10: Description Of Controls

    DESCRIPTION OF CONTROLS Selected Current Indicators Light when the switch under the indicator is pressed. Output Test Indicators Indicate normal or abnormal test con- Output Indicators ditions. Light to indicate the selected current (CUT, BLEND or COAG) is being delivered Defective Connection Indicators when the footswitch is depressed.
  • Page 11 Cord Hanger Main Body & Components Carrying Handle Ventilation Grill (rear) Power Cord Connector (Appliance Inlet) (front) Ventilation Fuse Holder Grills Ratings Plate Terminal TEST plug PATIENT plug (to endoscope) S-cord Check Lead S-P Cord (to Patient-plate) ACTIVE plug P-cord Patient-plate Pedal Footswitch...
  • Page 12: Construction

    C O N S T R U C T I O N Panel Panel Display Control Circuit Circuit Waveform Control Amplifier OsciIlator Generating Circuit Circuit Comparator Endoscope Foot- switch Active electrode When the footswitch is depressed, active current (I ) flows through the A-cord. Return current (I ) flowing through the patient plate to the P-cord connector is moni- Patient...
  • Page 13: Standard Set

    S T A N D A R D S E T Please check each item in the set against the list of standard components below. Contact Olympus if there are any missing or defective parts. Check Lead (MB-576) Electrosurgical Unit PSD-10...
  • Page 14: Preparation For Use

    5-1 Installation of PSD-10 Check that all safety precautions described in Section 1 – 3 are met. Place the PSD-10 on a stable surface in a horizontal position. Ventilation must not be obstructed. Carts and stands must be of adequate size and strength to support load.
  • Page 15: Connection Of Footswitch

    Red mark 5-4 Connection of S-P Cord Connect the S-P cord to the PATIENT connector on the PSD-10, with the red dot facing upward. Push the plug until it is locked in place. Red mark Connect the P-cord connector to the Patient-plate.
  • Page 16: Inspection Of Electrosurgical System

    INSPECTION ELECTROSURGICAL SYSTEM Power Feeding Turn ON the power switch. The power indicator and LED (Light-Emitting Diode) in- dicators on the front panel will light and the cooling fan will be heard. If the unit fails to turn on, inspect the power cord connection. Turn on power switch.
  • Page 17 WARNING light is glowing red. Re-connect the S-P cord correctly and repeat above procedure. (If the red WARNING light fails to go out, consult an Olympus service center.) Do not touch the Patient-plate; burns may occur.
  • Page 18 Clip combination with CD-3U/L or CD-4U/L. Be sure to perform the Output Check 1 for the PSD-10 to ensure the proper connection of the system. Connect the A-cord to the PSD-10 (ACTIVE connector) and to the electro- surgical accessory (A-cord plug on SD-handle).
  • Page 19 Checking the Output Adjustment Switches Press a given output adjustment switch (0.5 to 5.0): the indicator above the switch should light. Checking the Current Selector Switches Press a given current selector switch: the waveform pattern above the switch should light. Checking the Connection Monitor Connection between S-cord and endoscope Loosen the clamping ring on the S-cord connector.
  • Page 20: Endoscopic Electrosurgery

    ENDOSCOPIC ELECTROSURGERY This section outlines a general procedure for endoscopic electrosurgery. The endoscopist should evaluate the clinical factors involved and decide on the specific details of the procedure. 7-1 Precautions against Patient/Operator Burns To prevent accidental burns, attention must be paid to the following points. The patient should not come into contact with grounded metal objects (e.g., iron bar of bed, external surface of an appliance).
  • Page 21: Setup Of Patient-Plate

    7 - 2 S e t u p o f P a t i e n t - p l a t e Incomplete attachment of the Patient-plate may cause patient burns. Put gauze pads which are slightly larger than the Patient-plate and moistened with physiological saline or conductive paste (e.g., Conductive Gel #1103, 3M) on the Patient-plate.
  • Page 22: Electrosurgery

    7-3 Electrosurgery The type of output current and its intensity must be determined by refer- ence to related papers and based on one’s own experience. Power Feeding Turn ON the power switch. Selection of Current Press the desired current selector switch (CUT, BLEND or COAG). If none of the current selector switches are pressed, none of the current indicators above them will be lit;...
  • Page 23 Connection of A-Cord SD-handle Connect the A-cord to the PSD-10 and to the electrosurgical accessory. A-cord Only Olympus A-cord MA-255 is compatible with this equipment. As an added precaution, connect to the accessory as the final step before de- pressing the footswitch.
  • Page 24: After Operation

    7-4 After Operation Turn OFF the power switch. Disconnect the power plug from the hospital grade receptacle if the PSD-10 is not going to be sued for an extended period. Non-Controlled Document...
  • Page 25: Care And Storage

    Surfaces must be thoroughly dry before use. Wipe panel clean. 8 - 2 S t o r a g e Wind the power cord around the hanger on the rear of the PSD-10. Band Place the PSD-10 on a stable surface in the horizontal position.
  • Page 26: Maintenance

    9-1 Replacement of Fuse If the PSD-10 ceases to operate and the malfunction cannot be isolated, check the fuse holder on the rear of the PSD-10 after unplugging the power cord from the hospital grade receptacle. Be sure to unplug the power cord from the AC mains.
  • Page 27: Electrosurgical System Chart

    C L V - F 1 0 S-P Cord MB-575 Patient-plate MB-574 CLV-U20 Foot Release MB-332 Electrosurgical S-P Cord for Disposable (Disposable Patient-Plate) Unit PSD-10 SNARE Patient-plate MB-583 Patient-plate No. 1149 (3M) A-Cord MA-255 SD-Handle MA-254 Check Lead MB-576 Footswitch MB-399 Footswitch MB-585...
  • Page 28 San-Ei Building, 22-2, Nishi Shinjuku 1-chome, Shinjuku-ku, Tokyo, Japan (Premises/Goods delivery) Wendenstrasse 14-16, D-20097 Hamburg, Germany. (Letters) Postfach 10 49 08, D-20034 Hamburg, Germany. 4 Nevada Drive, Lake Success, N.Y. 11042-1179, U.S.A. KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United Kingdom BLK 211, Henderson Road No.