Appendix 1 Unpacking And Inspection; Appendix 2 Troubleshooting; Status Leds; Relay (Outputs) Troubleshooting - Xylem YSI IOEM Safety And Installation Manual

Input/output, relay, expansion modules
Table of Contents

Advertisement

Appendix 1 Unpacking and Inspection

Inspect the outside of the shipping carton for damage. If damage is detected, contact
the carrier immediately. Remove the instrument from the shipping container. Be
careful not to discard any parts or supplies. Confirm that all items on the packing list
are present. Inspect all assemblies and components for damage. Save the original
packing carton. Carriers typically require proof of damage due to mishandling.
The following components are included with the purchase of the IOEM and REM
instrument:
IOEM and REM instrument
006515 Flange Mounting Kit
006506 Desiccant Kit
605876 Instruction Manual
605230 Compression Plugs, 2 each
605229 Compression Plugs, 2 each
605804 Compression Plugs, 1 each
655384 Dual Male RJ-45 Cable Assembly
655383 Adapter, DB-9 to RJ-45
605880 AC Power Cord, 115 VAC U.S. Plug (AC version only)
605944UL Pluggable Socket Connector for AC power, 3 pole (AC Version only)
506939UL Pluggable Socket Connector for 12 VDC Power, 2 pole
605941UL Pluggable Socket Connector for RS-485, 5 pole
605943UL Pluggable Socket Connector for relay, 12 pole
605940UL Pluggable Socket Connector for Auxiliary inputs, 3 pole
If any parts are damaged or missing, contact YSI or your YSI representative im-
mediately.
IOEM/REM Safey and Installation Guide
Unpacking and Inspection
84

Appendix 2 Troubleshooting

Status leds

There are 2 status leds located on IOEM and REM devices. The leds are located left
of the DC power connector location at A on IOEM and REM devices
The PWR status LED is illuminated solid green when 12v power is applied. The COM
status LED has 3 states as described below:
1.
OK (RS485 communications active)
A. Short (100mS) flash every second. This is the normal and most common
state.
2.
Communication error
A. Fast blinking (on 200mS/off 200mS)
3.
Bootloader, when firmware is being uploaded
A. Slow blinking (on 500mS/off 500mS)
See IOEM reset to factory - page
See REM reset to factory - page

Relay (Outputs) Troubleshooting

- Disable control, alarm, and timer relays when troubleshooting relay
outputs.
- Verify IOEM and REM wiring and configuration.
Issue
possible cause
incorrect sensor configura-
relay does not energize
tion
incorrect timer configuration
30 second delay at power up
and/or sensor configuration
change
peripheral device is wired
incorrectly
IOEM/REM Safey and Installation Guide
Troubleshooting
possible solution
See Set Points, Controls,
and Alarm, 5X00 instru-
ment user manuals.
is correct relay configured?
See Timers, 5X00 instru-
ment user manuals.
See 30 seconds delay,
5X00 instrument user
manuals.
See Wire Relays - page
40 and Test Relays - page
34.
85