Configure 5X00 System For Ioem 4/20Ma Output; Configure 5X00 System For Ioem Pid Output Control Device - Xylem YSI IOEM Safety And Installation Manual

Input/output, relay, expansion modules
Table of Contents

Advertisement

Configuring IOEM/REM expansion modules

Configure 5X00 system for IOEM 4/20mA output

1.
Follow steps 1-6 of hardware installation-starting on page 25;
2.
Wire IOEM I/O for 4/20 output type- page 47;
3.
Verify all steps at Using AquaManager for IOEM configuration - page
73 are completed;
4.
Open AquaManager>open 5X00 node window>
Select 5X00 system tab to be configured;
5.
Configure 5X00 4/20mA output system fields;
6.
5X00 Systems Management PID control mode with 4/20mA output configuration
Notes:
-
must be selected to make 4/20mA fields live;
- Proper wiring of IOEM output channel is required in order for IOEM
4/20mA output device to work correctly.
- IOEM Source channels are only available at 5X00 system fields once
IOEM mode has been uploaded using AquaManager-page 73;
- IOEM Sources will only be available at IOEM and channel mode pull-
down menus that have been configured as a 4/20mA output type; For 5X00
system configuration field information see - AquaManager Help>Contents
tab>Working with Nodes>Systems management>Systems Management
fields table and link to specific sensor system being configured bookmarks;
IOEM/REM Safey and Installation Guide
;
76
Configuring IOEM/REM expansion modules
- 4/20mA output channels can only be assigned to 1 5X00 system at a
time. When selecting a 4/20mA output channel that is currently pending
or confirmed AquaManager displays the below 4/20mA Conflict Error
window.
conflict is resolved
- 4/20mA output values cannot be viewed at 5X00 instruments or at
AquaManager.
to upload system configuration to networked instrument;
7.
Test configured output functionality after reconfiguring - see Chapter 4
8.
5X00 instrument manuals Configuring>Sensor Setup>Service> Output test
system menu to verify 4/20mA output.
Verify 4/20mA signal. The 4/20mA output signal is not displayed at the
9.
Run Screen. Therefore, it is strongly recommended to verify the 4/20mA
signal at the IOEM or third party device.

Configure 5X00 system for IOEM PID output control device

1.
Follow steps 1-6 of hardware installation-starting on page 25;
2.
Wire IOEM I/O for 4mA-20mA output type- page 47;
3.
Verify all steps at Using AquaManager for IOEM configuration - page
73 are completed;
4.
Open AquaManager>open 5X00 node window>
Select 5X00 system tab to be configured;
5.
Configure 5X00 system fields
6.
- For 5X00 system configuration field information see - AquaManager
Help>Contents tab>Working with Nodes>Systems management>Systems
Management fields table and link to specific sensor system being con-
figured bookmarks;
IOEM/REM Safey and Installation Guide
will not be live until the 4/20mA assignment
;
77