4.3 Setting up the ventilator breathing circuit

1
Select hose system Leak valve system or expiration
valve system
1
Open the sliding cover on the underside of Carina.
3
2
Using a coin, turn the switch to your hose system
3
Close the sliding cover.
Leak valve hose system:
4
Connect the filter to socket.
5
Connect the ventilation hose: VentStar Carina LV
(MP00312) to filter.
The hose system is displayed on the status line of the
screen as follows: »LeakV«.
5
or:
Expiration valve hose system:
4
Connect the filter to socket.
5
Connect the ventilation hose: VentStar Carina EV
(MP00313) to filter
6
Connect the pilot line of the expiration valve to
pilot hose socket.
5
The hose system is displayed on the status line of the
screen as follows: »ExpV«.