Linearization - KERN HFC Service Manual

Electronic crane scales
Table of Contents

Advertisement

9.2

Linearization

Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a monitoring of test resources, you can
improve this by means of linearization.
Carrying out linearization is restricted to specialist staff possessing well
acquainted with the workings of balances.
The test weights to be used must be adapted to the balance's
specifications
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
After successful linearization you will have to carry out calibration
Preparation:
Call up linearization function "LinE0"; see chap. 8.2
How to carry out linearization:
Confirm „LinE0" by
"LoAd0" will be displayed. Make sure that no loads apart
from the lifting tackle are attached to the hook.
Wait for stability display, then press
As soon as "LoAd 1" is displayed attach first adjustment
weight (1/3 max).
Wait for stability display „STABLE", then press
As soon as "LoAd 2" is displayed attach second adjustment
weight (2/3 max).
HFC-SH-e-1510
.
.
.