MagCore Plus II Service Manual page 27

Hide thumbs Also See for Plus II:
Table of Contents

Advertisement

A. Test the Tip of the W2.
1. Please put the Tip Holder Set into the Well 2 of
the T-Rack.
2. Please add 1 ml water to the Sample Tube and
put it in the W1 of the T-Rack.
3. Please select the Test W2 Tip.
4. The test can be divided into two stages. The first
stage is to observe the smooth of taking the tip
out of the Tip Holder. Please check whether the
tip holder is taken out together with the tip. If
the Tip Holder is taken out together with the tip,
please reset the Tip Holder and try it again.
5. Please check whether the liquid levels of all tips
are consistent with each other. If the liquid levels
of all tips are not consistent with each other,
please check the O-Ring (Please refer to the
Replace O-Ring chapter for more information).
B. Test the Tip of the W3
1. Please put the Pipette Tip into Well 3 of the
T-Rack.
2. Please add 1 ml water to the Sample Tube and
put it in the W1 of the T-Rack.
3. Please select the Test W3 Tip.
4. Please check whether the liquid levels in all tips
are consistent with each other. If the liquid levels
are not consistent, please check the O-Ring
(Please refer to the Replace O-Ring chapter for
more information).