MagCore Plus II Operation Manual

MagCore Plus II Operation Manual

Hide thumbs Also See for Plus II:

Advertisement

Quick Links

CONTENTS
━ Operating Parameters . . . . . . . . . . . . . . . . . . . . . . 5
━ Operating Environment . . . . . . . . . . . . . . . . . . . . . 5
━ Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
━ Before Installation . . . . . . . . . . . . . . . . . . . . . . . . . 6
━ System Overview . . . . . . . . . . . . . . . . . . . . . . . . . 7
━ Install Reagent Cartridge . . . . . . . . . . . . . . . . . . . . . 8
━ Install Tube, Tip . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
━ Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
━ Installation Settings And Test . . . . . . . . . . . . . . . . . . . 9
━ Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
━ Emergency Stop . . . . . . . . . . . . . . . . . . . . . . . . . 14
━ Start Programs . . . . . . . . . . . . . . . . . . . . . . . . . . 15
━ Saved Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 16
━ Access to Saved Settings . . . . . . . . . . . . . . . . . . . . 16
━ Set and Select Users . . . . . . . . . . . . . . . . . . . . . . . 17
━ Set User . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
━ Select User Name . . . . . . . . . . . . . . . . . . . . . . . . 17
━ Scan Sample / Elution Tube . . . . . . . . . . . . . . . . . . 18
━ Scan Kit Lot . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
━ Pause and Abort . . . . . . . . . . . . . . . . . . . . . . . . . 19
━ Set Up the System Time . . . . . . . . . . . . . . . . . . . . 19
━ Set the Instrument Nickname . . . . . . . . . . . . . . . . . 20
━ Network Connection Settings . . . . . . . . . . . . . . . . . 20
━ LIMS Data Collection . . . . . . . . . . . . . . . . . . . . . . 21
1 . Get Data from USB Flash Drive . . . . . . . . . . . . . . . . 21
━ Remote monitoring APP . . . . . . . . . . . . . . . . . . . . 23
1 . Download the APP . . . . . . . . . . . . . . . . . . . . . . . 23
September 2022
EN-MagCore® Plus II
Operation Manual-1.0.0
MagCore® Plus II
Operation Manual
1
2 . Registration . . . . . . . . . . . . . . . . . . . . . . . . . . .23
2
4 . APP interfaces . . . . . . . . . . . . . . . . . . . . . . . . .24
4
5
━ Replace O-Ring . . . . . . . . . . . . . . . . . . . . . . . .27
━ Piston Test . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
━ Clean Piercing Needles . . . . . . . . . . . . . . . . . . . .30
━ Eject Tip . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
6
━ Heater Test . . . . . . . . . . . . . . . . . . . . . . . . . . .31
6
━ Alarm History . . . . . . . . . . . . . . . . . . . . . . . . . .32
━ System Status . . . . . . . . . . . . . . . . . . . . . . . . . .32
━ Program Update . . . . . . . . . . . . . . . . . . . . . . .33
━ Clean the Waste Box . . . . . . . . . . . . . . . . . . . . .33
━ Replace Fuse . . . . . . . . . . . . . . . . . . . . . . . . . .34
Document Revision History
12
15F, No.15, Qiaohe Rd,Zhonghe Dist, New Taipei City 235029, Taiwan
. . . . . . . . . . . . . . . . . . . . . .31
27
36

Advertisement

Table of Contents
loading

Summary of Contents for MagCore Plus II

  • Page 1: Table Of Contents

    MagCore® Plus II Operation Manual CONTENTS Safety Precautions 2 . Registration . . . . . . . . . . . . . . . . . . . . . . . . . . .23 Intended use 3 .
  • Page 2: Safety Precautions

    15. Object and Liquid Entry – Never push objects of any kind into this product as they Do not place metal objects inside the MagCore® System and wipe away all spilled may touch dangerous voltage points or short-out parts that could result in a fire or liquid to minimize the risk of system error and electric shock.
  • Page 3: Intended Use

    The MagCore® Nucleic Acid Extraction System is an automated nucleic acid purification system consisting of the MagCore® instrument, software, Kit (consumables and reagents). The MagCore® Nucleic Acid Extraction System is intended to be used by professional users for the purification of nucleic acids from biological samples for in vitro diagnostic purposes.
  • Page 4 Prevention of Cross-Contamination To minimize the risk of cross-contamination and to prevent contact with potentially infectious materials, always complete the processes involving sample preparation such pipetting of blood samples in a designated part of the benchtop or safety hood before starting PCR setup. To further minimize the risk of contamination, always carry out sample preparation, PCR setup, and the PCR run itself in separate rooms specially designated for each phase in the workflow.
  • Page 5: Ivd Symbols Reference

    MagCore® Plus II Reference Symbol IVD Symbols Reference Reference symbol Description of Symbol CE mark In Vitro Diagnostic Device Authorized representative in the European Community Manufacturer Content su cient for <n> tests Warning: electricity Instruction manual must be read before operating...
  • Page 6: Specification

    Specification Model Plus II System Method Cellulose Coated Magnetic Beads System Components 1. Pipetting Unit: Dispensing, transferring, X -Y two axis movements. 2. Electric Control: PLC module and Arm-based main board embedded in. 3. UV Light: Power 8W, life duration 11,000 hrs .
  • Page 7: Accessories

    MagCore® Plus II Accessories T-Rack X 1 Cartridge Rack X 1 Power Cable X 2 (Europe standard and U.S.A standard) Grease X 1 3mm O-Rings X 16 5mm O-Rings X 16 Empty Cartridge X 10 Sample Tube /Elution Tube X 20 Pipette Tip X 20 10.
  • Page 8: System Overview

    ━ System Overview Front Door Logo Indicator Touch Panel Operation Area Waste Box USB Port Scanner Port Main Power Switch Ethernet Port Power Outlet Back Service Lid www.rbcbioscience.com...
  • Page 9: Install Reagent Cartridge

    MagCore® Plus II ━ Install Reagent Cartridge Please insert the front end of the cartridge into the space below the fixing plate of the Cartridge Rack. Pull cartridge rack outward (toward user), take it out and place consumables accordingly, please insert cartridges...
  • Page 10: Installation

    ━ Installation ━ Installation Settings And Test Step 1. Step 1. Please Set Up the Local Time. 1. Open the Front Door 2. Turn on the machine Cushioning Material Accessory Box Please refer to System Status chapter for the information about the operation.
  • Page 11: Installation

    Installation For Restricted Pers MagCore® Plus II ━ Installation Open the Front Door Remove cushioning material and take out accessory box. Romove cushioning material Take out accessory box Accessory Box Accessory Box Cushioning Material...
  • Page 12 Personnel Use Remove the brackets and cable ties then pull out the touch panel. Remove the two brackets Close the front door. Move the rack module outward manually, then cut off and remove the two cable ties Switch on the power, the machine will initialize automatically. Pull out the touch panel and tighten screw after touch panel is erected.
  • Page 13: Touch Panel And User Interface

    MagCore® Plus II Touch Panel and User Interface MagCore® Plus II provides 7-inches full-color touch panel and easy-to-use user interface. Operator only needs to pay attention to the panel to know the current status, ending time and test report of the machine. The figure below shows the function list of the operation screen and the description of the status icons.
  • Page 14: Operation Before Starting The Instrument

    Operation Before Starting the Instrument Precautions When Opening the Front Door Plunged Upward from The Front Door Will For the sake of safety, the front the Lower Edge of the Rise Automatically door will not open to the top. Front Door Please push the front door up to raise it to the top when needed.
  • Page 15: Emergency Stop

    MagCore® Plus II ━ Emergency Stop Time Remaining Estimated finish time Program Code 0 : 0 Please read the security information to correctly stop the machine when an accident takes place. Pause Current Status 01/31/2019 13:15:47 When an emergency situation occurs and the machine is operating, please press the stop button or open the front door to stop the machine immediately.
  • Page 16: Start Programs

    ━ Start Programs Select DNase Treatment. (Only for RNA kit.) Please pretreat the sample according to the instructions of the user manual of the MagCore® Kit and put consumables into the machine. Select Sample Volume. Press Start to go to the next step.
  • Page 17: Saved Settings

    MagCore® Plus II ━ Saved Settings ━ Access to Saved Settings After loading the Tips and Cartridges: If you need to save a customized setting, you can save it here and use the predetermined parameters for later uses. Press Saved Protocols on the start page After selecting the parameters, the confirmation of your settings will be displayed.
  • Page 18: Set And Select Users

    ━ Set and Select Users Instrument can help you record the User Name as well as the Sample, Elution Tubes and Kit Lot No. used in the extraction for LIMS integration data. Scan Sample Scan Elution Tube Select User Jemes Scan Kit LOT Back Next...
  • Page 19: Scan Sample / Elution Tube

    MagCore® Plus II ━ Scan Sample / Elution Tube ━ Scan Kit Lot Scan Sample Scan Elution Tube Scan Sample Scan Elution Tube Select User Select User Jemes Scan Kit LOT Jemes Scan Kit LOT Back Next Back Next Select...
  • Page 20: Pause And Abort

    ━ Pause and Abort ━ Set Up the System Time You can pause or abort the current extraction while the program is in progress. Press the Pause button to pause at any time, and the machine will immediately suspend operation. This button is used as an emergency stop button.
  • Page 21: Set The Instrument Nickname

    MagCore® Plus II ━ Network Connection Settings ━ Set the Instrument Nickname You can give the instrument an easy-to-remember The network connection is used for the remote nickname for remote APP identification. monitoring APP and LIMS data transmission. A.For the remote monitoring APP function, please...
  • Page 22: Lims Data Collection

    B.Connection fail ━ LIMS Data Collection 1. Please confirm the software version with your distributor. 2. If the software version is correct, please refer to There are two ways to get the data : the following content and contact your IT staff. 1.
  • Page 23: Access To Information By Network Connection

    MagCore® Plus II 2. Access to information by B.Alarm history: usbdisk / disk_b_1 / AlarmLog / datalog Network Connection Please plug the USB Flash Drive into the instrument USB port, select "Report" and click "Save Data". Obtaining data via internet connection. We provide FTP protocol as a means of data transmission.
  • Page 24: Remote Monitoring App

    ━ Remote monitoring APP 3. Log in the APP and Pair the Instrument Click "+", then scan the instrument QR code to pair. 1. Download the APP The APP supports both iOS and Android systems. You can scan the QR Code below or search i-moldx in the store to download and install.
  • Page 25: App Interfaces

    MagCore® Plus II 4. APP interfaces A.Click the instrument to check the current status. Instrument type Start / End time Serial number Selected parameters Extraction step, pause, abort, alarm messages...
  • Page 26 B.APP Setting: Sort and Unpair the instrument Click 1. Click to move and sort the position of the instrument. 2. Click to un-pair. C.APP Setting: Add Administrator User Allow other users to use the APP without re-registering their account. Login account must be Admin in order to change or manage users. Select User Management Currently there is only one user, who Click...
  • Page 27 MagCore® Plus II Enter the account (e-mail account is suggested) Enter the password Enter the e-mail account Whether the account is the Admin Whether the account is suspended Select User Management Complete...
  • Page 28: Easy Maintenance

    Easy Maintenance Easy Maintenance includes the maintenance items which need to be done by users. These items include sterilization, clearing, simple test and error message report, etc. Users must perform these checkups periodically to ensure the system status is always good.
  • Page 29 MagCore® Plus II A. Test the Tip of the W2. B. Test the Tip of the W3 1. Please put the Tip Holder Set into the Well 2 of 1. Please put the Pipette Tip into Well 3 of the the T-Rack.
  • Page 30 C. Test the Tip of the W4 1. Please put the 200μl SP Tip into Well 4 of T-Rack. 2. Please add 1 ml water to the Sample Tube and put it in W1 of the T-Rack. 3. Please select the Test W4 Tip. 4.
  • Page 31: Clean Piercing Needles

    MagCore® Plus II ━ Clean Piercing Needles ━ Eject Tip If the system is stopped from power shortage or other abnormal conditions, and the tips are still on the pistons, you have to press Reject Tip to remove the tips from the pistons.
  • Page 32: Heater Test

    ━ Heater Test ━ UV Sterilization Please Select Heater Test Button. Please Select the Sterilization Button. Please put the thermometer into the Heat Block Well 1 and press Next. Please select the sterilization time and press Start button. Warning! The UV lamp is dangerous. Please do not look straight at the inner of the machine when it is operating.
  • Page 33: Alarm History

    MagCore® Plus II ━ Alarm History ━ System Status If the machine malfunctions or the following You can check the system status and set up the messages appear, the engineers may ask you to system time in the menu. download the error records for them to diagnose the machine in advance.
  • Page 34: Program Update

    ━ Program Update ━ Clean the Waste Box You can use the USB to update the system. Please copy the downloaded program to the root directory of the USB Flash Drive.(Not provided) Please dump the waste when the program is Then insert the USB Flash Drive into the USB Port completed, and wash the waste box every time.
  • Page 35: Replace Fuse

    MagCore® Plus II ━ Replace Fuse 250V 10A Please use the specified type of fuse to minimize the risk of burning.
  • Page 36 Automated Nucleic Acid Extractor Model : Plus II Input Raitng :100-240V~ 50/60Hz, Max . 4 . 5 AMax, 450W RBC Bioscience Corp . 15F . , No . 1 5, Qiaohe Rd . , Zhonghe Dist . , New Taipei City 235029, Taiwan...
  • Page 37 MagCore® Plus II Document Revision History Date Changes Updated the following sections: General Considerations; September 2022 Document Revision History EN-MagCore® Plus II Operation Manual-1.0.0 Update to comply with the requirements of the IVD Regulation (EU) 2017/746.
  • Page 38 15F, No.15, Qiaohe Rd,Zhonghe Dist, New Taipei City 235029, Taiwan TEL : +886 2 8912 1200 FAX : +86 2 8912 1300...