Functioning Principle, Technical Data
3.2
Technical Data
Standard measurement ranges, typical values (±5 %) for extended measurement range in brackets
Measurement range, z axis
Start of measurement range (SMR)
Reference distance, midrange
End of measurement range (EMR)
1
Resolution z axis
1
s
Linearity z axis (3 )
Measurement range x axis
Aperture angle of the laser line
Resolution x axis
1
s
Linearity x axis (3 )
0.4 %
Resolution profile parameters
Profile frequency (profiles/second)
Measurement rate (measuring points / s)
Light source
Laser class (EN 60825-1)
Laser switch-off
Permissible extraneous light (fluorescent lamp)
Protection class
Operating temperature
Storage temperature
Electromagnetic compatibility
(EMC)
Vibration (acc. to IEC 60068-2-6)
Shock (acc. to IEC 60068-2-29)
Sensor dimensions (without cable)
Sensor weight (without cable)
Sensor connecting cable, standard length
Controller dimensions (without connectors)
Controller weight (without cable)
Output analog
preassigned: x and z axis,
alternatively: profile parameters
Interfaces
Serial (measurements and control commands)
Switching signals
4
Digital inputs
4
Digital outputs
Power supply
All specified data applies to metallic mattfinished surfaces.
1) For standard measurement range, slight displacement of the measuring field possible (depends on sensor)
2) 1024 points per profile with extended measurement range and 128,000 points/s (optional 256,000 points/s) only.
3) Only applies to values preprocessed in the controller, e.g. width, height, glue bead area, edge detection, groove width, angle etc.
4) Preassigned, other functions possible, e.g. encoder.
5) The data apply to the sensor.
X9751109-B060097MSC
Model
0.04 %
0.2 %
at SMR
at EMR
256 / 512 / 1024
at SMR
at EMR
up to 1000 profiles/s (optional up to 4000 profiles/s)
Laser diode 655 nm, 15 mW (optional 50 mW)
5
5
(LxWxH) mm
3
Sync-In, Remote laser ON/OFF, User mode, Encoder
LLT28x0-25
25 (55) mm
62.5 (50) mm
75 (82.5) mm
87.5 (105) mm
10 µm
50 µm
13 (23) mm
18 (41) mm
30 °
2
, optional 64 / 128 points/profile
60 µm
80 µm
0.05 %
up to 256,000 measuring points / s
Class 2M (optional: 3B)
Remote input and key switch
10,000 lx
Sensor IP 64, controller IP40
0...50 °C
-20...70 °C
EN 61000-6-3 Spurious emission
EN 61000-6-2 Resistance to disturbance
2 g / 20 ... 500 Hz
15 g / 6 ms
109 x 64 x 44
350 g
2 m
L x W x H 278 x 187 x 107 mm
3.5 kg
± 10 V (16 Bit, up to 150 kHz)
R
= 50 Ohm, I
l
max
3x IEEE 1394 („FireWire"), 400 MBit/s,
DCAM 1.30 Standard
RS232 / RS422; 115,200 Baud
Sync-Out, error, user mode (2x)
24 VDC ± 15 %; 0.5 A
MICRO-EPSILON
LLT28x0-100
100 (245) mm
145 (115) mm
195 (235) mm
245 (360) mm
40 µm
200 µm
30 (50) mm
50 (140) mm
120 µm
200 µm
125 x 64 x 44
400 g
= 5 mA