MODULE SPECIFIC SYNTAX
The
Module
Specific
Multiplexer/SPDT Switch is required in the use of the OPEN and
CLOSE commands. It will also appear in data output by the Master
in response to the PDATAOUT and PSETUP commands.
The Module Specific Syntax for the 1260-37 module is as follows:
<module address> .<channel>[ ;<module address> <channel>]
where <module address> is the switch card address.
<channel number> is the relay to be closed to connect
an input to the output.
Note that Channels remain closed until opened by an OPEN or
RESET command, VXI hard or soft reset, or power-off.
The <module address> used here is not the VXIbus defined
logical address of the 1260 Series Master. It is particular to
the 1260 Series and describes the switching module in
relation to the Master. This address corresponds to the
binary value of the switch setting of SW1 on the switching
module PCB.
The range of values for <channel> is:
Multiplexer/Scanner:
Chapter 3
Syntax
for
the
1260-37
NOTE:
One-wire
Two-wire
Four-wire
SPDT Switch
Module Specific Syntax 3-1
Signal
00-48
00-47
00-23.
100-139