Table 2-1, 1260-37 Multiplexer/Scanner Circuit Jumper Installation

8(1X6)
4(1X6)
2-Wire
4-Wire
W2A,B
W3A,B
W4A,B
W5A,B
W6A,B
W8A,B
W9A,B
W10A,B
W11A,B

Analog Bus

EADS North America Defense
Test and Services, Inc. © 1994
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
An X indicates a jumper is to be installed. An (X) indicates the jumper
is optional, depending on whether access to the analog bus is
required. A blank indicates no jumper is to be installed.
4(1x12)
2(1X12)
2-Wire
4-Wire
X
X
X
X
X
X
X
X
In
two
of
the
Scanner/Multiplexer circuit may be configured to access the analog
bus (refer to Figure 3-1). The analog bus allows expansion for the
configuration of larger scanner/multiplexers than the module may
achieve alone. This is accomplished by providing access to a
common bus channel which may be daisy chained to other
multiplexer modules via the front panel
To connect the module to the analog bus, install jumpers W2A and
W2B on the motherboard PCB.
2(1X24)
1(1X24)
2-Wire
4-Wire
X
X
X
X
X
X
X
X
X
X
X
X
above
configurations,
Installation Instructions 2-3
1(1X48)
1(1X96)
2-Wire
1-Wire
(X)
(X)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
the