8. WIRING AND CONNECTION

■8.1 Wiring
1) Refer to the table below to obtain a suitable shielded cable.
Model number
□□
KH15-J-
terminal box
(small)
□□
KH15-K-
terminal box
(large)
2) Process the end of the cable in accordance with Figure 8-1 and 8-2 below for each model number.
□□
< KH15-J
(small terminal box) >
Strip 60 mm of the cable sheath and 7 to 9 mm of the core coat. Provide shielding at the bottom
end of the sheath. Cable that fits the terminal block is AWG24 to 12 (0.2 to 3.3 mm
□□
< KH15-K
(large terminal box) >
Strip 80 mm of the cable sheath, and then strip the core coat to fit the length of the M3 crimp
terminal to be used. Mount the crimp terminal to the wiring terminal. Provide shielding at the
bottom end of the sheath.
Terminal opening size
10a
(standard)
G3/8
10b
15a
15b
G1/2
(standard)
15c
< Table 8-1 >
< Figure 8-1 >
< Figure 8-2 >
Gasket inner dia.
(mm)
φ7
φ8
φ9
φ10
φ11
(60)
80
14
Suitable cable dia.
(mm)
φ6 to 7
φ7 to 8
φ8 to 9
φ9 to 10
φ10 to 11
2
).
7 to 9
For M3
Y-KH15-007A Rev.3