SECTION 15—SEAT

Adjusting/Replacing Telescoping Front Rigging Support

ASBA Seats
NOTE: For this procedure, refer to FIGURE 15.11
NOTE: The two telescoping front rigging supports can be positioned at different depths depending
on the need of the user.
1. Remove the two mounting screws, spacers and locknuts that secure the telescoping
front rigging support to the seat frame.
2. Perform one of the following:
• Slide existing telescoping front rigging support to one of three depth positions.
• Remove existing telescoping front rigging.
3. Secure telescoping front rigging at desired depth with existing two mounting screws,
spacers, and locknuts. Torque locknuts to 75 in-lbs ± 20%.
FIGURE 15.11 Adjusting/Replacing Telescoping Front Rigging Support - ASBA Seats
Adjustable ASBA Seats
NOTE: For this procedure, refer to FIGURE 15.12 on page 99.
1. Remove the two cap screws, spacers and threaded blocks securing the telescoping
front tube to the side rail.
2. Perform one of the following:
• Slide existing telescoping front rigging support to one of five depth positions.
• Remove existing telescoping front rigging.
3. Secure the telescoping front tube to the side rail at the desired depth with the existing
two cap screws, spacers and threaded blocks. Torque cap screws to 79 in-lbs.
4. Repeat STEPS 1 to 3 on the opposite side, if desired.