Immunity
IEC 60601
Test
Test Level
ESD
±6 kV Contact
IEC 61000-
±8 kV Air
4-2
+/- 6 kV
Contact
+/- 8 kV Air
EFT
±2 kV Mains
IEC 61000-
±1 kV I/Os
4-4
Surge
±1 kV
IEC 61000-
Differential
4-5
±2 kV
Common
Voltage
>95% Dip for
Dips/
0.5 Cycle
Dropout
60% Dip for
|IEC 61000-
5 Cycles
4-11
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Power
3 A/m
Frequency
50/60 Hz
Magnetic
Field IEC
61000-4-8
14
Electromagnetic
Compliance
Environment –
Level
Guidance
+/- 6 kV
Floors should be
Contact
wood, concrete or
+/- 8 kV Air
ceramic tile.
Note: Reset
If floors are synthetic,
may occur at
the r/h should be at
6 kV
least 30%.
Test not
Mains power quality
applicable
should be that of a
typical commercial or
hospital environment.
Test not
Mains power quality
applicable
should be that of a
typical commercial or
hospital environment.
Test not
Mains power quality
should be that of a
applicable
typical commercial or
hospital environment.
If the user of the
SurfLink Programmer
requires continued
operation during
power mains
interruptions, it
is recommended
that the SurfLink
Programmer be
powered from an
uninterruptible power
supply or battery.
3 A/m
Power frequency
magnetic fields
should be that of a
typical commercial or
hospital environment.
Guidance and Manufacturer's Declaration -
Electromagnetic Immunity
The SurfLink Programmer is intended for use in the electromagnetic
environment specified below. The customer or the user of the SurfLink
Programmer should assure that it is used in such an environment.
Immunity
IEC 60601
Test
Test Level
Recommended Separation Distance
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
Radiated RF
80 MHz
IEC 61000-4-3
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Electromagnetic
Compliance
Environment –
Level
Guidance
Portable and mobile
RF communications
equipment should
be used no
closer to any part
of the SurfLink
Programmer,
including
cables, than the
recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
3 Vrms
d = 1.2√P
150 kHz to
80 MHz
3 V/m
d = 1.2√P 80 MHz
80 MHz to
to 800 MHz
2.5 GHz
d = 2.3√P 800 MHz
to 2.5 GHz
15
Need help?
Do you have a question about the SurfLink and is the answer not in the manual?