Safety instructions General Assure yourself prior and after the use of the RESCUE 230 that the unit is in safe and usable condition (cables integrity, battery status). RESCUE 230 is not intended for use in areas of highly inflammable anesthetics or other inflammable substances, especially in oxigenous areas.
Page 4
INTRODUCTION Thank you for choosing the RESCUE 230. Please read this Operator’s Manual carefully and thoroughly before using the RESCUE 230. This Manual contains instructions on how to operate and maintain the RESCUE 230. It is very important that you fully understand all the necessary instructions discussed in this manual so as to act quickly in an emergency.
The warranty period of this device is within one year after the date of purchase. When the device malfunctions during the warranty period (RESCUE 230 subjected to operating conditions within the specified limits), it will be repaired free of charge at our service centers.
Page 6
Also include the name of model, product serial number, date of purchase, name of sales representative, Customer. information There are no user-serviceable parts in the RESCUE 230. The user should not attempt any repair beyond what we recommend in the maintenance section of this manual.
Page 7
The following safety messages are used to emphasize the safety practices that must be observed and followed during the operation of the RESCUE 230. The operator must follow the instructions in all the Danger, Warning, Caution, and Notice messages found throughout this Operator’s Manual In the event that the device is damaged due to misuse or negligence by a user, the manufacturer or the authorized representatives shall not be responsible for the damage or loss to the product.
Page 8
Device Operation Guidelines General Guidelines Do not operate or store the device in conditions that are beyond the following specified limits. Operating Conditions Temperature 0 °C to 50 °C Humidity 5 % to 95 % (non-condensing) Storage Conditions Temperature -20 °C to 70 °C Humidity 5 % to 95 % (non-condensing)
Page 9
Keep the device away from dusty environments. There are no user serviceable parts inside the RESCUE 230. Only authorized service personnel should open the device for repairs. During recharging, when the device is connected and disconnected from the mains power connector, hold the plug and not the electrical cord.
Cleaning and Maintenance After each use, clean the RESCUE 230 using a soft, damp cloth moistened with any of the following solvents: Soap and water 70% solution isopropyl alcohol Chlorine bleach and water mixture (30 ml bleach/liter of water) Ammonia-based cleaners Hydrogen peroxide Do not immerse any part of the RESCUE 230 in fluids.
40 milliseconds of the occurrence of a QRS peak. The RESCUE 230 may also be used to do ECG monitoring. ECG monitoring only is done by connecting the custom designed, 4 leads ECG monitoring cable assembly from ELPRO S.r.l.
Intended Purposes Indications Asynchronous defibrillation – the shock delivery is not synchronized with the QRS peak In asynchronous defibrillation, the RESCUE 230 is indicated for use on patients with the following symptoms a) Unconsciousness b) Absence of normal breathing and c) Lack of detectable pulse.
Page 13
Operational front panel controls and indicators SYNCRO MODE SELECTION INTERNAL DISCHARGE ENERGY SELECTION LCDisplay PRINT START/STOP PAPER FEED RELEASE SHOCK ENERGY CHARGE BATTERIE CHARGING INDICATOR ON/OFF BUTTON SpO2 ECG CABLE CONNECTOR CONNECTOR PRINTER...
Page 14
Key SYNC The sync mode is displayed on the screen. Pressing this key will discharge the energy internally. (If no discharge is actuated, after 30 sec the RESCUE 230 will discharge internally the energy). Key DISCH Used to select the energy level desired. The corresponding light...
When the charge is finished this light will switch off. Charging Light When RESCUE 230 is switched on, the light may be on even if the (on the ON/OFF batteries are fully charged (the device is in backup mode, using the...
Page 16
Verify the pads contact with the patient chest and add more conductive gel. ARMING RESCUE 230 is charging the capacitor to the energy level selected. Indicates that the capacitor is charged and the device is ready for defibrillation. In Sync mode the shock will be released only when a valid READY ECG wave R peak is detected.
Display organization When the 4 leads ECG patient cable is disconnected, the display will show only one ECG trace (II) input from the defibrillation pads. Connecting the 4 leads ECG patient cable the display will show 3+3 ECG traces: I,II,III or aVR, aVL, aVF based on the user selection (see menu functions).
AC POWER SUPPLY CONNECTOR Input Connectors Used for the defibrillation pads. Defibrillation Pads input For the disposable pads please order also the adapting (left side down) cable. Used for the SpO2 sensor (optional). The RESCUE230 will SpO2 input automatically recognize sensor will modify...
Device Preparation Unpacking Carefully inspect the packing container and the device for any damage that might have been sustained during shipping. Check the shipping list to ensure that the unit comes with the complete accessories. It is important to have all the necessary accessories all the time. Make sure that you have the complete accessories during unpacking.
The ARMING message will be displayed and the charging sound will start. When the charge ends the READY message will be displayed and on the SHOCK key the red light will go on. This means that the RESCUE 230 is ready for defibrillation.
4. To release the defibrillation shock, press both push buttons on the standard pads or press the SHOCK key if using disposable pads. The RESCUE 230 will discharge internally after a 30 sec wait time. If the defibrillation is not required, press the DSCH (F4) key to discharge internally.
The battery charge light should go on. When the batteries are fully charged the light will go off. To see the battery charge status, switch on the RESCUE 230 with at least one pad disconnected. The battery charge status will be displayed. If it shows less than 75% than please connect the AC charger.
Page 23
Pressing the F1 key the ECG monitoring will start. Set-up menu The set-up menu enables to preset the start up (at switch on) parameters of the RESCUE 230. To access the set-up menu, press the MENU key twice to reach the second menu, then press F2 key (SET-UP).
Scope of Delivery • R230-001 RESCUE 230 console • R230-002 Standard defubrillation pads • R230-003 Four leads ECG patient cable • R230-004 AC charger power supply • R230-005 User manual • R230-006 Carrying case • R230-007 Thermal paper roll Optional accessories...
Page 25
Do not use the reusable internal paddles with an energy setting higher than 50 Joules. Progetti cannot be hold responsible for a non correct use of the internal reusable paddles or for an energy setting higher than 50 Joules. User manual RESCUE 230 _ver EN 2.2...
Page 26
• Input: 100 ~ 240V AC 50/60Hz max. 1A • Output: 12V DC, 2.5A Battery pack • 12V battery Ni-Cd (internal rechargeable) • charging time 3 to 4 hours • capacity : 100 shocks (battery fully charged) User manual RESCUE 230 _ver EN 2.2...
Page 27
Biphasic waveforms The following waveforms are the RESCUE 230 output for a maximum of 230 Joule: User manual RESCUE 230 _ver EN 2.2...
Rescue 230 dovrebbe assicurarsi che esso venga usato in tale ambiente. The equipment Rescue 230 is intended for use in the electromagnetic environment specified below. The customer or the user of Rescue 230 should assure that it is used in such an environment.
Rescue 230 deve garantire che esso venga usato in tale ambiente. The equipment Rescue 230 is intended for use in the electromagnetic environment specified below. The customer or the user of Rescue 230 should assure that it is used in such an environment.
Page 30
IEC 61000-4-8 commercial or hospital environment. è la tensione di rete in c.a. prima dell’applicazione del livello di prova. Nota_e is the a.c. mains voltage prior to application of the test level. User manual RESCUE 230 _ver EN 2.2...
Page 31
RF fissi, si dovrebbe considerare un’indagine elettromagnetica del sito. Se l’intensità di campo misurata nel luogo in cui si usa l’apparecchio Rescue 230, supera il livello di conformità applicabile di cui sopra, si dovrebbe porre sotto osservazione il funzionamento normale dell’apparecchio Rescue 230.
Page 32
The equipment Rescue 230 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the equipment Rescue 230 can help prevent electromagnetic interference...
CE DECLARATION OF CONFORMITY ELPRO s.r.l. declares that the device RESCUE 230 with all its accessories Is conform to the quality assurance system - normative 93/42/CEE. Certified body : M. Masini institute N° 0068 – Con formity certificate number: NR. 0068/QCO-DM/031-2009 Rev. 5 del 19/05/12...