Negative pressure wound therapy

Advertisement

Troubleshooting
1. Low Pressure/Leak
Low
WHAT YOU
Pressure/
SEE/HEAR
Leak
BLINK
WHAT TO DO
1
Push up
on bottom
of canister.
2
If you see:
SOLID LIGHT
Low Pressure/Leak
is fixed.
2. Canister Full/Blockage - with clamp
WHAT YOU
Canister
Full/
SEE/HEAR
Blockage
BLINK
WHAT TO DO
First, look at the canister to see if it is full. If the canister is full, change
the canister (see "Changing the Canister" on front page). If the canister
is not full, proceed with the steps below.
Make sure
1
clamp is
OPEN.
3
Look for kinks in
tubing. Unkink
if found.
5
6
Turn OFF.
Clamp tubing and remove
canister
Canister" on front page)
7
Replace with new canister,
reattach tubing and open clamp
(see "Changing the Canister" on front page)
Wait
If you see:
9
1 minute
SOLID LIGHT
Canister Full/Blockage
is fixed.
Cardinal Health
Waukegan, IL 60085 USA
Rev C 2019-05 • cardinalhealth.com
68-08-0013
© 2019 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO, CATALYST and TWIST N' CONNECT are trademarks or registered trademarks
of Cardinal Health and may be registered in the US and/or in other countries. All other marks are the property of their respective owners. US Pat. 7,608,066
+
+
BEEP
BLINK
Wait 30 seconds
If you see/hear:
+
+
BLINK
BEEP
BLINK
Call Cardinal Health
at 1.866.484.6798
+
+
BEEP
BLINK
Wait
If alert
2
30 seconds
continues,
go to step 3.
Wait
4
If alert
30 seconds
stays on, go
to step 5.
(see "Changing the
.
8
Turn ON.
.
If you see/hear:
+
+
BLINK
BEEP
BLINK
Call Cardinal Health
at 1.866.484.6798
68-1132
REF
2. Canister Full/Blockage - without clamp
WHAT YOU
Canister
Full/
SEE/HEAR
Blockage
BLINK
WHAT TO DO
First, look at the canister to see if it is full. If the canister is full, change
the canister (see "Changing the Canister" on front page). If the canister
is not full, proceed with the steps below.
1
2
If you see:
If you see/hear:
SOLID LIGHT
BLINK
Canister Full/Blockage
is fixed.
NOTE: A false Canister Full/Blockage alert can occur if drainage moves into the canister while the
CATALYST™ is lying on its front (the button side) or if it is turned upside down. If this occurs, you must
change the canister to resume device operation. Always keep the CATALYST™ upright.
3. Battery Low
WHAT YOU
SEE/HEAR
BLINK
WHAT TO DO
1
Plug A.C. Power Adapter into outlet.
2
If you see:
SOLID LIGHT
Low battery fixed.
4. Intermittent Mode
WHAT YOU
SEE/HEAR
WHAT TO DO
CATALYST™ makes more noise every 5 minutes when in Intermittent Mode.
No troubleshooting needed. For more information, contact
your healthcare professional.
+
+
BEEP
BLINK
Turn ON.
Wait
30 seconds
+
+
BEEP
BLINK
Call Cardinal Health at 1.866.484.6798
+
CHIRP
Wait 30 seconds
If you see/hear:
+
OFF Button Flashing
+ Chirp
Call Cardinal Health
at 1.866.484.6798
BLINK