16 . EXTERNAL INPUT / OUTPUT CONNECTOR

16 - 1 External Input / Output Connector A
(1) Details of External Input / Output Connector A Signals
WARNING
・ DO NOT connect any circuit other than SELV (DC+24V) (Safety Extra Low Voltage) to the External
Input / Output Connector A of the CONTROLLER. This will cause I / O board damage in the
CONTROLLER.
・ Do not supply over voltage or over current into the input / output circuit. Always install a LOAD
(resistor) to the output circuit to eliminate the chance of damage to the CONTROLLER.
CAUTION
External Input / Output Connector A DOES NOT use Pins No. 9 and No. 22. If pin No. 9 and No. 22
are connected, the CONTROLLER will be damage.
Table. 4
Pin
Code
Function
No.
1
COM_1
External Power
source for
External input
2
DIR_IN
Rotating
Direction Setting
3
CNT_IN
Count Pulse
Signal for Setting
Motor Speed
4
RESET
Error Release
5
SEL1
Speed Point
Select 1
6
RUN
Rotating
7
DIR_OUT
Rotating
Direction
Input /
Output
Input
DC 0V or DC+24V
Input
OFF (Open) : FWD.
ON (Closed) : REV.
Input
OFF (Open) → ON (Closed) One pulse will increase or decrease
Input
ON (Closed) → OFF (Open) Error Code can be released and the
Input
OFF (Open)
ON (Closed)
Output OFF (Open) : Stop
ON (Closed) : Rotating
Output OFF (Open) : FWD.
ON (Closed) : REV.
Description
Power source to be used for External
Inputs Signals. (not included / prepared
by the end-user)
Controls the rotational direction of the
motor spindle.
Setting parameter
, can start with
reverse rotation. (Refer to P93 " 18 - 4
Selection of External Motor Start
Signal Control Mode " section.)
-1
100min
(rpm) in Spindle Speed depending
on parameter
setting. (Refer to P92
" 18 - 4 ⑤
Selection of External
Speed Control Mode " section.)
system restarted by toggling this signal
OFF and ON.
Error will not be released until cause of
the error has been removed.
Speed Point (
Selected by SEL0 and SEL1 signal
combination. P70 " 15 - 2 - 2 (4) ③ Set by
the Speed Point Signal Table 3 " section.
Need to set parameter
(Refer to P92 " 18 - 4 ⑤
External Speed Control Mode " section.)
Output shows that the motor is rotating.
Output shows the direction of the Motor is
rotating.
) can be
.
Selection of