new brunswick Innova 43 Operation Manual page 85

Incubator shakers
Table of Contents

Advertisement

12.6
Report Request Commands
Code
RI
Report
Software
Version
RP
Report
Parameter
List (ASCII
text header)
Report
RV
Measured
Values
Report
RS
Setpoint
Values
12.7
Set/Get Date & Time
Code
Meaning
=D
Set Date
& Time
?D
Get Date
& Time
New Brunswick Scientific Co., Inc.
Meaning
RI<CR>
Return data format:
<Arg1><Tab><Arg2><CR><LF>
where Arg1=Display module software version, and
Arg2=Control module software version
RP<CR>
Return data format:
<Arg1><Tab><Arg2><Tab><Arg3>...<Arg6><CR><LF>
where Arg1= Shaker Speed ,
Arg2= Temperature ,
Arg3= % Relative Humidity ,
Arg4= Reserved ,
Arg5= Grow Lamp Status , and
Arg6= UV Lamp Status
RV<CR>
RS<CR>
Return data format:
<Arg1><Tab><Arg2><Tab><Arg3>...<Arg6><CR><LF>
where Arg1= Shaker Speed,
Arg2=Temperature,
Arg3=% Relative Humidity,
Arg4=(Reserved, reads 0.0),
Arg5=Grow Lamp Status, and
Arg6=UV Lamp Status
=D<Space><Arg1><Space><Arg2><Space><Arg3><Space>...
<Arg7><CR>
?D<CR>
Return data format:
<Arg1><Tab><Arg2><Tab><Arg3><Tab>...<Arg7><CR><LF>
where Arg1=Hours (range 0-23),
Arg2=Minutes (range 0-59),
Arg3=Seconds (range 0-59),
Arg4=Year (range 00-99),
Arg5=Month (range 01-12),
Arg6=Date (range 1-31),
Arg7=Day (range 1-7; 1=Monday, 2=Tuesday...7=Sunday)
Format
Format
User s Guide
85

Advertisement

Table of Contents
loading

Related Products for new brunswick Innova 43

This manual is also suitable for:

Innova 43r

Table of Contents