new brunswick Innova 43 Operation Manual page 84

Incubator shakers
Table of Contents

Advertisement

84
Code
PR
PS
PW
PW<Space>25.0<Space>150<Space>0.0<Space>2<Space>30<Space>0
<Space>1<CR>
Temperature setpoint 25.0º
Agitation setpoint 150 rpm
Innova 43 & 43R Incubator Shakers
Meaning
Read
PR<Space><Arg1><Space><Arg2><CR>
Profile
where Arg1= Profile number (range 1 to 4) and
step in
Arg2=Step number (range 1 to 15)
memory
Return data format:
<Arg1><Space><Arg2><Space><Arg3><Space>...<CR><LF>
where Arg1=Profile number (range 1 to 4),
Arg2=Step number,
Arg3=Temperature setpoint,
Arg4=Agitation setpoint,
Arg5=(Reserved, reads 0.0),
Arg6=Step hours,
Arg7=Step minutes,
Arg8=UV lamp, and
Arg9=Grow lamp
Profile
PS<CR>
Stop/
With no other parameters, any profile currently running will be
Start
stopped.
(stops or
starts a
PS<Space><Arg1><Space><Arg2><CR>
program
where Arg1=Profile number (range 1 to 4), and
profile)
Arg2=Step number (range 1 to 15).
NOTE: If Arg2 is not included, the profile will start in Step 1.
Write
PW<Space><Arg1><Space><Arg2><Space><Arg3><Space>
Profile
...<CR>
step to
where Arg1=Profile number (range 1 to 4),
memory
Arg2=Step number (range 1 to 15),
Arg3=Temperature setpoint,
Arg4=Agitation setpoint,
Arg5=(Reserved); Enter 0.0,
Arg6=Step hours (range 0 to 99),
Arg7=Step minutes (range 0 to 59),
Arg8=UV lamp (0 = off, 1 = on), and
Arg9=Grow lamp (0 = off, 1 = on)
M1320-0050
Format
PW
Step Time 2 hours 30 minutes
UV Lamp OFF
GRO Lamp ON
User s Guide

Advertisement

Table of Contents
loading

Related Products for new brunswick Innova 43

This manual is also suitable for:

Innova 43r

Table of Contents