new brunswick Innova 43 Operation Manual page 83

Incubator shakers
Table of Contents

Advertisement

12.4
Set Commands
Code
CL
CS
CT
CU
12.5
Profile Control Commands
Code
PC
PM
New Brunswick Scientific Co., Inc.
Meaning
Command Grow
Lamp
Command Speed
Command
Temperature
Command UV
Lamp
Meaning
Clear Profile
(clears profile
step)
Profile Mode
(returns status of
current profile
run/hold mode)
CL<Space><Arg1><CR>
where Arg1= Set Grow Lamp state (0 = off, 1 = on)
CS<Space><Arg1><CR>
where Arg1= Agitation setpoint (####)
CT<Space><Arg1><CR>
where Arg1= Temperature setpoint (###.#)
CU<Space><Arg1><CR>
where Arg1= Set UV Lamp state (0 = off, 1 = on)
PC<Space><Arg1><Space><Arg2><CR>
where Arg1= Profile number (range 1 to 4) and
Arg2=Step number (range 1 to 15)
NOTE: If Arg2 is not included, all 15 steps of the
selected profile will be cleared.
PM<CR>
There are no parameters to this command.
Return data format:
<Arg1><Space><Arg2><Space><Arg3><CR>
where Arg1=Run/hold status (0 = end, 1 = run),
Arg2=Profile number and
Arg3=Step number
NOTE: If a profile has not been selected to run, a 0
will be returned for Profile number and Step number.
...continued...