new brunswick Innova 43 Operation Manual

Incubator shakers

Advertisement

Quick Links

Incubator Shakers
MANUAL NO: M1320-0050
Revision E
May 28, 2008
NEW BRUNSWICK SCIENTIFIC CO., INC.
BOX 4005 44 TALMADGE ROAD EDISON, NJ 08818-4005
Telephone: 1-732-287-1200
1-800-631-5417
Fax: 732-287-4222 Telex: 4753012 NBSCO
Internet: http://www.nbsc.com E-mail: bioinfo@nbsc.com

Advertisement

Table of Contents
loading

Summary of Contents for new brunswick Innova 43

  • Page 1 Incubator Shakers MANUAL NO: M1320-0050 Revision E May 28, 2008 NEW BRUNSWICK SCIENTIFIC CO., INC. BOX 4005 44 TALMADGE ROAD EDISON, NJ 08818-4005 Telephone: 1-732-287-1200 1-800-631-5417 Fax: 732-287-4222 Telex: 4753012 NBSCO Internet: http://www.nbsc.com E-mail: bioinfo@nbsc.com...
  • Page 2 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 3 INTERNATIONAL OFFICES: BELGIUM GERMANY CHINA THE NETHERLANDS FRANCE UNITED KINGDOM New Brunswick Scientific Co., Inc. User s Guide...
  • Page 4 Do not use this equipment in a hazardous atmosphere or with hazardous materials for which the equipment was not designed. New Brunswick Scientific Co., Inc. (NBS) is not responsible for any damage to this equipment that may result from the use of an accessory not manufactured by NBS.
  • Page 5 Reproduction, adaptation, or translation without prior written permission from New Brunswick Scientific is prohibited. Disclaimer Notice New Brunswick Scientific Co., Inc. reserves the right to change information in this document without notice. Updates to information in this document reflect our commitment to continuing product development and improvement.
  • Page 6 W A R R A N T Y Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 7: Table Of Contents

    TABLE OF CONTENTS OVERVIEW........................11 INSPECTION & UNPACKING OF EQUIPMENT ............13 PREPARING THE LOCATION..................14 INSTALLATION........................ 17 FEATURES ......................... 19 GETTING STARTED ......................26 New Brunswick Scientific Co., Inc. User s Guide...
  • Page 8 Running a Program....................50 7.7.1 Calculating the Offset Value................. 51 7.7.2 Setting the Offset....................51 MAINTENANCE........................ 54 SERVICE & ACCESSORIES ................... 56 9.11.1 Platforms....................... 67 9.11.2 Flask Clamps for Platforms .................. 68 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 9 Optional Utility Basket ..................70 9.11.10 Optional Remote Alarm ..................71 DRAWINGS, GRAPHS & TABLES ................72 SPECIFICATIONS......................77 11.1.1 CE Declaration of Conformity................79 APPENDIX: REMOTE PROGRAMMING ............... 80 APPENDIX: PRODUCT RETURNS................87 INDEX..........................89 New Brunswick Scientific Co., Inc. User s Guide...
  • Page 10 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 11: Overview

    Continuously: In a timed mode: Via the shaker s programmable controller: Via computer through an RS-232 interface. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 12 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 13: Inspection & Unpacking Of Equipment

    Inspection of Boxes Packing List Verification Unpacking of Equipment CRUSH WARNING! Do not attempt to lift the Innova 43/43R by hand. Always use a lifter or other suitable equipment when raising or handling the unit. Out of Box Concerns Customer Satisfaction Form 6300, Warranty Registration New Brunswick Scientific Co., Inc.
  • Page 14: Preparing The Location

    NOTE: The Innova 43/43R has casters and leveling feet. When you need to move the unit across the floor, make sure the feet are raised enough so that the casters can roll freely without scraping the floor with the feet. At all other times, make sure the feet are down, to avoid any unintended rolling.
  • Page 15 Space Requirements see Figure 1 Figure 1: Innova 43/43R Front View DEPTH: 32 inches (81 cm) including front WIDTH: panel protrusions 45.5 inches (115.6 cm) HEIGHT: HANDLE TO 40.6 inches OPEN LID (103 cm) with feet NOTE: Open lid adds 23.5 inches...
  • Page 16 Being very careful not to hit the control knob or button Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 17: Installation

    Leveling the Shaker NOTE: The Innova 43/43R has casters and leveling feet. When you need to move the unit across the floor, make sure the feet are raised enough so that the casters can roll freely without scraping the floor with the feet. At all other times, make sure the feet are down, to avoid any unintended rolling.
  • Page 18 Section 6 see Section 7 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 19: Features

    Whenever power to the shaker may be a hazard (during cleaning, maintenance or service work), be sure to disconnect the power cord from the electrical outlet. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 20 TEMPERATURE OFFSET ºC PARAMETERS ACTUAL PARAMETER VALUES SCREEN NAME DAY & 24-HOUR TIME DISP We 13 : 44 PROGRAM RUNNING AUDIBLE ALARM HEATER ON MUTED LID OPEN PARAMETERS LOCKED Changing Screens Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 21 Shaking speed, in revolutions/minute ºC Chamber temperature, in degrees Celsius Programmed time remaining, in hours Relative Humidity, in percent. Status of Ultraviolet germicidal lamp Status of Photosynthetic growth lamps Optional Display Icons New Brunswick Scientific Co., Inc. User s Guide...
  • Page 22 This icon appears when a user-defined Program is running. See Section 7.5.5. This icon appears to the right of ºC if the Temperature Offset feature is being used. See Section 7.7. Alarms Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 23 Indicates unit is powering up (both at normal power-up and POWER after power interruption); will flash until the Control Knob is moved. Indicates when timed run is completed. see Figure 1 AUTO Spill Pan/Water Reservoir New Brunswick Scientific Co., Inc. User s Guide...
  • Page 24 Section 6.7 (see Section 6.6 5.10 Heater NOTE: Operating the shaker for sustained periods at temperatures above 60°C may result in a shorter fan lifespan. 5.11 Refrigeration (43R Only) Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 25 5.12 Service Accessibility by an authorized service technician New Brunswick Scientific Co., Inc. User s Guide...
  • Page 26: Getting Started

    Figure 5 below, which also shows the spill cover that you must install Figure 5: Installing Platform & Spill Cover PLATFORM SCREWS PLATFORM SPILL COVER BEARING HOUSING Installation of Clamps see Section 9.11.2 see Figure 6 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 27 Figure 6: Double Girdle Clamp Installation CLAMP UPPER GIRDLE MOUNTING HOLES (5) WITH GIRDLE TUBES PLATFORM LOWER CLAMP BODY GIRDLE WITH (LEGS AND BASE) GIRDLE TUBES Figure 6a: Clamp Fastener Electrical Connections ON/OFF New Brunswick Scientific Co., Inc. User s Guide...
  • Page 28: Adjustable Foot

    Optional Gassing Manifold Kit This option is factory-installed. see Figures 8 & 9 on the following page NOTE: Gas supply should be regulated and it must never exceed 15 PSIG. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 29: Gas Manifold (Optional)

    Never exceed 15 PSIG inlet pressure to the manifold. Make sure free lengths of tubing are secured by the hook and loop straps to keep them from contact with flasks or other moving parts. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 30: Germicidal Uv Lamps (Optional)

    FRONT OF SHAKER Optional Photosynthetic Lamps see Figure 11 on the following page The recommended operating temperature when using this option is 15 - 37ºC; the maximum operating temperature is 70ºC. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 31: Photosynthetic Lamps (Optional)

    Figure 11: Photosynthetic Lamps (optional) PHOTOSYNTHETIC LAMP ASSEMBLY Optional Humidity Monitor Figure 12a: Humidity Monitor (optional) HUMIDITY MONITOR New Brunswick Scientific Co., Inc. User s Guide...
  • Page 32 Figure 12b: RH Sensor Operating Range & Error Graph Operating Range and Humidity Error +/- 5% Error INTERMITTENT OPERATION +/- 3% Error +/- 5% Error Temperature (C) Figure 12c: RH Sensor Connection BOARD inside the Service M1282-7002 M1282-7012 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 33 Figure 12d: Humidity Sensor Mounting INSIDE WALL OF CABINET CAUTION! Operating the Innova 43/43R at temperatures above 60ºC can result in permanent damage to the sensor. See Section 6.9 for instructions. Filling the Water Reservoir NOTE: As you add water, do not allow the water to splash or flow into the recessed middle of the pan, where the bearing housing assembly is mounted.
  • Page 34 6.10 Draining the Water Reservoir Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 35: Operation

    Safety Precautions Product Returns CAUTION! To prevent damage to the shaker and its contents, never run the shaker without a platform. Opening the Lid New Brunswick Scientific Co., Inc. User s Guide...
  • Page 36: Display Screen

    NOTE: UV (UV Germicidal lamp) and GRO (photosynthetic growth lamps) will appear in this screen with the word NONE if your shaker is not equipped with these optional features. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 37: Changing Display Parameter

    Figure 13: Changing Display Parameter ºC DISP We 13:44 see Figure 14 Figure 14: Changed Display Parameter ºC DISP We 13:44 To view a setpoint ºC New Brunswick Scientific Co., Inc. User s Guide...
  • Page 38 If you highlight an item, change it, but do not save your selection, after a few seconds the screen will revert to its previous setting. To move out of this screen DISP DISP SUMM SUMM Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 39: Summary Screen

    To change setpoints in this screen NOTE: If you highlight an item, change it, but do not save your selection, after a few seconds the screen will revert to its previous setting. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 40: Setup Screen

    To change the day NOTE: If you highlight an item, change it, but do not save your selection, after a few seconds the screen will revert to its previous setting. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 41: Lamps Screen

    To lock the settings Lock see Section 5.4 Lock To mute the audible alarm Mute Mute To move out of this screen LAMP LAMP 7.4.4 Lamps Screen see Figure 18 Chamber UV Decont Growth New Brunswick Scientific Co., Inc. User s Guide...
  • Page 42 If you highlight an item, change it, but do not save your selection, after a few seconds the screen will revert to its previous setting. To move out of this screen LAMP LAMP Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 43: Rs232 Screen

    Display is at Revision D and information F/W Rev : the FirmWare only) Control Board at Revision H. COMM Th 16:21 To change the Communication Mode Table 5: Communication Mode, on the following page New Brunswick Scientific Co., Inc. User s Guide...
  • Page 44: Calibrate Screen

    To move out of this screen COMM COMM 7.4.6 Calibrate Screen see Figure 20 on the following page for details, see Sections 7.7, Temperature Offset Calibration, and 0, Using Calspeed Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 45: Programs Screen

    Programs Screen see Figure 21 Figure 21: Programs Screen PROGRAMS Program 1 : Program 2 : Program 3 : Program 4 : PROG Th 16:23 Programming the Shaker 7.5.1 Timer Only DISP SUMM New Brunswick Scientific Co., Inc. User s Guide...
  • Page 46: Programmed Steps

    Edit Off Off Figure 21: Programs Screen PROGRAMS Program 1 : Program 2 : Program 3 : Program 4 : PROG Th 16:23 7.5.3 Creating a Program see Figure 22 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 47 If your shaker is not equipped with these optional features, UV and GRO will appear with the word Off in this screen but will not be programmable. duration see Figure 23 20.0 ºC ºC see Figure 23 see Figure 23 New Brunswick Scientific Co., Inc. User s Guide...
  • Page 48 Cancel NOTE: The time entered for each step is for that step only; it is not cumulative (i.e., the elapsed time from the start of the program). see Figure 24 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 49: Editing A Program

    Step 3 PRG1 - STEP Step number Time 03:00 12:00 ºC 40.0 Save Cancel To save the entire program Save Process Running Saving Profile PROG 7.5.4 Editing a Program Use the function Edit New Brunswick Scientific Co., Inc. User s Guide...
  • Page 50: Running A Program

    Program 3 : Program 4 : PROG Fr 08:31 RUN icon To stop a program: Muting the Audible Alarm see Table 4 Mute To mute the alarm, Temperature Offset Calibration Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 51: Calculating The Offset Value

    Figure 20, repeated below DISP SUMM NOTE: If you highlight an item, change it, but do not save your selection, after a few seconds the screen will revert to its previous setting. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 52 Figure 20 above Calspeed NOTE: Prior to calibrating the speed, make sure that the platform is properly secured to the subplatform, and any flasks present are secured. Power Interruption automatic restart Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 53 New Brunswick Scientific Co., Inc. User s Guide...
  • Page 54: Maintenance

    Routine Maintenance see Section 8.2 below for more details Cleaning External & Internal Surfaces If Biohazard decontamination is required, see Section 8.3 below. Biohazard Decontamination Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 55 Personnel involved in the cleanup of any spill should wear gloves, safety glasses, and a laboratory coat or gown during the cleanup process. Respiratory protection should be considered for spills where aerosolization is suspected. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 56: Service & Accessories

    Incorrect speed calibration: recalibrate shaking speed. and/or no speed Defective main board: call for service. indication. Defective motor: call for service. Drive belt out of alignment or worn: call for service..continued... Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 57 Incorrect temperature to C on display). indication. Defective RTD assembly: call for service. Defective main board: call for service. Incorrect relative Defective humidity probe or broken contact: call for service. humidity % ...continued... New Brunswick Scientific Co., Inc. User s Guide...
  • Page 58 Check ballast voltage. Check cable connections. Check switch settings. UV lamp mode has been changed by RS-232 command/ computer interface. Product Returns A form for this purpose is provided in Section 13 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 59: Fuse Replacement

    Table 6 below Table 6: Fuses Fuse No. Function Type Rating Heater 6.3 A GRO Lamps (photosynthetic) 3.15 A ® Slo Blo UV Lamp (germicidal) 3.15 A Compressor 10 A optional equipment New Brunswick Scientific Co., Inc. User s Guide...
  • Page 60: Replacing Cabinet Lights

    NOTE: Cabinet lights are connected in series. When one burns out, both lights will be off. It is therefore prudent to replace both bulbs at the same time. WARNINGS NOTE Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 61: Replacing Photosynthetic Lamps

    Figure 27: Replacing Cabinet Lights Replacing Optional Photosynthetic Lamps WARNING! Before removing or installing the photosynthetic GRO lamp, make sure power is disconnected from the unit. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 62 Figure 28: Replacing Photosynthetic Lamps Adjusting Lid Tension WARNING! The following procedures are provided for your information only. Do not attempt to perform these service interventions yourself unless you are an authorized service technician. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 63: Germicidal (Uv) Lamp Replacement

    (3) Before removal, make sure the lamp has fully cooled. WARNINGS WARNINGS Figure 29: Germicidal (UV) Lamp Replacement ® NOTE that wires, tie wraps and Permagum are not shown above. New Brunswick Scientific Co., Inc. User s Guide...
  • Page 64 Belt Replacement or Adjustment To gain access to the drive belt in order to prevent it from dropping against your fingers or causing you to lose your grip. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 65: Drive Assembly

    Figure 30: Drive Assembly SPILL DEFLECTOR BEARING DRIVE HOUSING ASSEMBLY To replace the drive belt New Brunswick Scientific Co., Inc. User s Guide...
  • Page 66: Belt Replacement & Adjustment

    Figure 31: Belt Replacement & Adjustment DRIVE BELT PULLEY BEARING HOUSING FOUR MOTOR FIXING BOLTS (loosen for belt tensioning) MOTOR If an adjustment is required: Reassembly: see Section Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 67: Platforms

    Table 8: Available Platforms Description NBS Part No. Universal Platform M1250-9920 50 mL Erlenmeyer flask Dedicated Platform M1191-9908 125 mL Erlenmeyer flask Dedicated Platform M1191-9909 250 mL Erlenmeyer flask Dedicated Platform M1191-9910 ...continued... New Brunswick Scientific Co., Inc. User s Guide...
  • Page 68: Flask Clamps For Platforms

    50 ml Erlenmeyer Flask M1190-9000 125 ml Erlenmeyer Flask M1190-9001 250 ml Erlenmeyer Flask M1190-9002 500 ml Erlenmeyer Flask M1190-9003 1 L Erlenmeyer Flask ACE-1000S 2 L Erlenmeyer Flask ACE-2000S ...continued... Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 69: Replacement Clamp Hardware Kits

    (38 mm) long. Angled Test Tube Rack Spacer Bar* for use with TTR-210 to accommodate test tubes racks that are less than 5 in. TTR-215 (13 mm) wide. * Universal Platform Required New Brunswick Scientific Co., Inc. User s Guide...
  • Page 70: Optional Gas Manifold Kit

    Description NBS Part Number Humidity Monitor (for use with spill pan as water M1320-0600 reservoir) 9.11.9 Optional Utility Basket Table 16: Utility Basket Description NBS Part Number Utility Basket M1320-0700 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 71: Optional Remote Alarm

    Figure 32: Utility Basket (optional) 9.11.10 Optional Remote Alarm New Brunswick Scientific Co., Inc. User s Guide...
  • Page 72: Drawings, Graphs & Tables

    & & 10.1 Schematics Figure 33: Control Schematics NOTE: For a larger version of this drawing, contact your sales or service representative. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 73 Figure 34: Refrigeration Schematic 10.2 Load & Speed Graphs New Brunswick Scientific Co., Inc. User s Guide...
  • Page 74 Figure 36: Load & Speed for 2-Inch Orbit Shakers (20% Full) 125, 250, 500 mL 2.8L <20 20 to 30 30 to 36 36 to 40 40 to 50 50 to 60 >60 Platform and Load (lb) Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 75 10.3 List of Drawings Figure Description Page Innova 43/43R Front View Adjustable Foot Front Panel (Detail) Control Panel (Right Side) Installing Platform & Spill Cover Double Girdle Clamp Installation Clamp Fastener Rear Panel Gas Manifold (optional) Gas Fitting (with optional Gas Manifold)
  • Page 76 Fuses Replacement Parts Available Platforms Platform Flask Capacities Flask Clamps Racks & Trays Gas Manifold Kit Photosynthetic Growth Lamps Germidical [UV] Lamp Humidity Monitor Utility Basket COM Parameters for HyperTerminal Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 77: Specifications

    34 pounds (15.5 kg) Innova 43 & 43R Incubator Shaker SHAKING 20-500 rpm with 1-inch (2.5 cm) stroke or 20-300 rpm with 2-inch (5.1 cm) Speed stroke Control Accuracy 1 rpm Indication Displayed in 1 rpm increments 1-inch (2.5 cm) or 2-inch (5.1 cm)
  • Page 78 Innova 43 & 43R Incubator Shaker CHAMBER 24 inches deep X 34.5 inches wide X 19 inches high from top of platform (61 cm deep X 87.6 cm wide X 48 cm high from top of platform) DIMENSIONS Foot-activated lid opening.
  • Page 79: Ce Declaration Of Conformity

    11.1.1 CE Declaration of Conformity New Brunswick Scientific Co., Inc. User s Guide...
  • Page 80: Appendix: Remote Programming

    Enter the setting on your shaker Baud Rate (see Step 2 above) Data Bits None Parity Stop Bits Flow Control None see Sample Test below Caps Lock Sample Test: Enter Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 81 Slave shaker will always send these four characters: . Before OK<CR><LF> starting any communications, be sure to clear the computer s receive buffer. Command Space Arg1 Space Argn Terminator New Brunswick Scientific Co., Inc. User s Guide...
  • Page 82 Read Profile Step in Memory Profile Stop/Start Write Profile Step to Memory Report Software Version Report Parameter List Report Setpoint Values Report Measured Values Set Date and Time Get Date and Time Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 83 Arg1=Run/hold status (0 = end, 1 = run), Arg2=Profile number and Arg3=Step number NOTE: If a profile has not been selected to run, a 0 will be returned for Profile number and Step number..continued... New Brunswick Scientific Co., Inc. User s Guide...
  • Page 84 Arg8=UV lamp (0 = off, 1 = on), and Arg9=Grow lamp (0 = off, 1 = on) PW<Space>25.0<Space>150<Space>0.0<Space>2<Space>30<Space>0 <Space>1<CR> Temperature setpoint 25.0º Step Time 2 hours 30 minutes Agitation setpoint 150 rpm UV Lamp OFF GRO Lamp ON Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 85 Get Date ?D<CR> & Time Return data format: <Arg1><Tab><Arg2><Tab><Arg3><Tab>...<Arg7><CR><LF> where Arg1=Hours (range 0-23), Arg2=Minutes (range 0-59), Arg3=Seconds (range 0-59), Arg4=Year (range 00-99), Arg5=Month (range 01-12), Arg6=Date (range 1-31), Arg7=Day (range 1-7; 1=Monday, 2=Tuesday...7=Sunday) New Brunswick Scientific Co., Inc. User s Guide...
  • Page 86 =D<Space><01><Space><30><Space><00><Space><04><Space><07> <Space><04><Space><07><CR> Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 87: Appendix: Product Returns

    13.1 Return Procedure 13.2 Return Authorization and Decontamination Certificate New Brunswick Scientific Co., Inc. User s Guide...
  • Page 88 I certify that the equipment described above has been thoroughly cleaned and decontaminated of all chemical, biological and radioactive contaminants and also certify that the returned unit is safe for unprotected human contact. Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 89: Index

    New Brunswick Scientific Co., Inc. User s Guide...
  • Page 90 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...
  • Page 91 New Brunswick Scientific Co., Inc. User s Guide...
  • Page 92 Innova 43 & 43R Incubator Shakers M1320-0050 User s Guide...

This manual is also suitable for:

Innova 43r

Table of Contents