RWP Weighing Indicator Technical Manual(v1.16)
 Enter the function by pressing
Currently adjustment
 If want to change by using the keys
select the required setting
 Enter the selected setting by pressing
display will be shown.
 Load the calibration mass weight on the platform and wait
few seconds for display stability.
 After the stable indicator on press
shown.
After the calibration the display will start a self test. Remove the load from
platform during the test. Display will come to weighing mode automatically.
If display will be shown any error or incorrect value, repeat the procedure again.
Linear Calibration
The linearity deviation caused by the performance of the weighing unit. The
digital linearization function can reduce the linearity deviation using weighing
points during the zero and capacity. Up to three weighing points can be specified.
 Press
 Enter the function by pressing
 Make sure there are no loads on the platform and wait few seconds for
stable indicator on.
 Enter the function by pressing
linear
continuous until display will be shown.
key, display will be shown
,
,
key ,
key , display will be
key , display will be shown
key , display will be shown
05. 000
keys to
load
pass
linear
Load 0
Load 1
- 22 -