Table 7-4: Device Status - Honeywell SmartLine STIW400 User Manual

Wireless temperature and universal i/o transmitters
Table of Contents

Advertisement

Transmitter display
OoS
SNSR ERR
'U' Status
LOW BAT
NO RADIO
The following status messages have multiple meanings. Refer to OneWireless UI Device Status for exact cause.
'F' Status
'F' Status
'F' Status
'F' Status
The following statuses are displayed only in OneWireless UI Device Status.
'U' Status
Revision 1

Table 7-4: Device status

OneWireless User
Interface display
OOS
Sensor Error
Calibration Error
Low Battery
Radio Interprocessor
Comm Error
Input Failure
NVM Fault*
Program Memory
Fault*
RAM Fault*
Excess Zero
Calibration
SmartLine Wireless Temperature and Universal I/O User's Manual
Definition
All channels are out of
service.
Sensor can not access
meter body A/D
converter.
Calibration Data Invalid
or could not be read.
Battery Voltage
Critically Low
Radio Board is not
accessible.
Input Error
Startup diagnostics
detected defect in
Sensor Non-Volatile
Memory
Startup diagnostics
detected defect in
Sensor Read Only
Memory
Startup diagnostics
detected defect in
Processor Random
Access Memory
The selected zero
offset or the lower
calibration trim point is
beyond 5% of the lower
end of the
characterized range of
the device.
7. Operation
7.2. Transmitter PV display
What to do
Restore mode to Auto in
OneWireless User Interface.
Check connection between
electronics module and meter
body. If still doesn't work,
replace sensor. See page 60.
Use Cal Clear, or User
Calibrate.
Replace batteries as soon as
possible. See page 61.
Restart both the radio and
sensor. If condition persists,
replace electronics module.
See page 60.
Possible meter body sensor
failure.
Replace electronics module.
See page 60.
Replace electronics module.
See page
60.
Replace electronics module.
See page 60.
Clear Calibration
47

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smartline stuw750Smartline stuw751