To Acquire the Calibration Service

RIGOL does not recommend manual calibration by users. If calibration is required,
please contact RIGOL customer service department or the local distributor.

Recommended Test Devices

It is recommended that you use the test devices listed in the table below or other test
devices whose performance specifications satisfy the "Performance Requirement"
listed in the table below for calibration.
Table 1-1 Recommended Test Devices
Device
Digital Multimeter
Test Lead
Short Circuitor

Test Notices

For optimum performance, all procedures should comply with the following
recommendations.
1)
Make sure the environment temperature is stable and between 18℃ and 28℃.
Ideally, the calibration should be performed at 23℃±1℃.
2)
Make sure the environment relative humidity is less than 80%.
3)
Make sure that the instrument has been warmed up for at least 30 minutes
before the calibration.
4)
Keep the test cables as short as possible and make sure that the impedance of
the cable is consistent with the requirement.
1-2
Performance Requirement
1
6
/
2
Connect the output terminals of
DP800 and the input terminals of
the multimeter
Short-circuit the positive terminal
(+) of the channel output
terminals and positive terminal
(S+) of the Sense terminals;
short-circuit the negative terminal
(-) of the channel output terminals
and negative terminal (S-) of the
Sense terminals
Chapter 1 Calibration Notices
Recommended
RIGOL DM3068
__
__
DP800 Calibration Guide