Pannoramic DESK 1.18.1 User's Guide
Hot surface! Wait for the lamp surface to cool down. Keep a cool-down
time of approximately 15 minutes.
When changing the lamp, avoid touching the replacement bulb with
the bare fingers!
1. Exit control software.
2. Switch off the power supply and disconnect the power cable.
3. Open the cover (see section 5.3.1 Removing the protective cover).
4. Hold the lamp mount (2) with your hand.
5. Remove the knurled screws (1) from the heat sink body (6) completely.
6. Turn the lamp mount counter-clockwise (for approximately 45 degrees) until the two
clamps (3) are aligned to the hole (5) on the heat sink body.
7. Remove the lamp mount with the lamp.
8. Remove the defective lamp (4) from the lamp mount.
9. Insert a new bulb into its mount.
Do not touch the bulb with your bare fingers! Use a dry, clean, lint-free
cloth or gloves when inserting the bulb into the mount.
10. Taking care that the two clamps (3) on the lamp mount are correctly aligned to the hole (5)
on the heat sink body, carefully push the lamp mount into the heat sink as far as you can.
11. Turn the lamp mount (2) clockwise (for approximately 45 degrees) aligning its bores with
the ones on the heat sink body.
12. Lock the lamp mount (2) with the knurled screws (1).
13. Reattach the cover (see section 5.3.1 Removing the protective cover).
14. Connect power plug and turn power supply on.
15. Launch control software.
May 29, 2014 - Rev. 1