Pannoramic DESK 1.18.1 User's Guide
Slide is often positioned incorrectly
Guiding fault in transport system
Digitization process is extremely slow
Extended focusing option is activated
Z-stack option is activated
Image is out of focus
Focus range limit is activated
Focus range setting is incorrect
All individual images show a shading effect
Algorithm failed to find empty image fields
May 29, 2014 - Rev. 1
Contact 3DHISTECH Service for support.
Deactivate the Extended focusing option if not required.
Deactivate the Z-stack option if not required.
Deactivate the Limiting focus range option.
Change focus range setting.
Capture a reference image with a blank slide (see section
Compensation).