Pannoramic DESK 1.18.1 User's Guide
The cover slip and the label area should be on the top, with the label
area pointing towards the slide insert door.
To avoid jamming or breaking slides, ensure that the slide is not tilted
and that it rests perfectly horizontally on the specimen holder.
Always make sure that the slide is correctly inserted on the specimen
holder. Incorrect slide insertion may break the glass. The edge of the
broken area of the slide is sharp and its removal can lead to injury and
contagion! Broken slide can be removed only by a well-trained person!
5. Close the fixing lever (2) by pushing it to the right so that it touches and therefore fixes the
slide.
6. Close the slide insert door (1) by pushing it to the right.
When closing the slide insert door, fingers may be pinched during
operation.
Never open the front door unless prompted accordingly by software.
When you open the door at any other time, make sure not to reach with
your hands into the instrument to avoid the risk of crushing your fingers.
7. When you close the door, barcode detection and preview image creation is started
automatically.
The preview image is displayed in the preview area of the control software.
May 29, 2014 – Rev. 1
Note:
Warning!
Caution!
Warning!
3DHISTECH Ltd.