Page 1
Keysight 42942A Terminal Adapter, 40 Hz to 110 MHz Operation and Service Manual...
Page 2
No part of this manual may be repro- nished under a license and may be ing elsewhere in the EULA. Keysight duced in any form or by any means used or copied only in accordance...
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the 42942A has been checked mechanically and electrically. The shipment should contain...
Follow the steps below to connect the 42942A to the Impedance Analyzer. Step 1. Join the 42942A terminal adapter to the test connectors on the front panel of the Impedance Analyzer by gently matching the four BNC connectors and fastening screws of the adapter with the test connectors and accessory mounting holes of the instrument until they come into complete contact.
Serial Number Keysight Technologies uses a ten-character serial number that is stamped on the serial number plate attached to the bottom of the four-terminal pair connection block. The first two characters shows the made-in country, the next three digits are assigned as the serial prefix, and the last five digits are the suffix.
Page 8
Installation Guide Serial Number Keysight 42942A Terminal Adapter...
Overview Product Overview 2 Overview Product Overview The 42942A is an adapter equipped with a 7-mm coaxial connector that connects a test fixture furnished with the same type of connector to the Impedance Analyzer. Figure 2-1 Product Overview Keysight 42942A Terminal Adapter...
Overview Functions Functions Figure 2-2 Figure 2-3 show the names of the 42942A’s parts and contents of the option 700, respectively. Figure 2-2 42942A Parts Table 2-1 42942A Function NAME FUNCTION OUTPUT Port Contact for the test fixture furnished with a 7-mm coaxial connector.
Page 11
FUNCTION OPEN An open device (OPEN, 0 S) used for the adapter setup. SHORT A short device (SHORT, 0 Ω) used for the adapter setup. LOAD A load device (50 Ω) used for the adapter setup. Keysight 42942A Terminal Adapter...
Never give any mechanical shock to the upper face with the 7-mm connector’s mouth. Also, do not place the 42942A with the upper face down on a hard surface. If the ambient temperature deviates by more than 5 ˚C from the temperature in which the adapter setup was done, it is recommended that you perform the adapter setup again in the ambient temperature.
Page 14
7. Connect the OPEN standard to the 7-mm OUTPUT port connector. Never store the OPEN standard loosely in a box or drawer because this may result in damage to the mating plane surfaces. Do not drop the OPEN standard or give it any strong mechanical shock. Keysight 42942A Terminal Adapter...
Page 15
OUTPUT port connector. Never store the SHORT standard loosely in a box or drawer because this may result in damage to the mating plane surfaces. Do not drop the SHORT standard or give it any strong mechanical shock. Keysight 42942A Terminal Adapter...
Page 16
Short [Done]. 12.Remove the SHORT standard from the 7-mm OUTPUT port connector of the 42942A. Then connect the LOAD standard to the 7-mm port. Never store the LOAD standard loosely in a box or drawer because this may result in damage to the mating plane surfaces.
Page 17
Impedance Analyzer Setup Figure 3-4 Connecting LOAD Standard 13.Press the Load [-] key to start the load data measurement. When the load data measurement is completed, the softkey label changes to Load [Done]. 14.Press the Save Impedance key. Keysight 42942A Terminal Adapter...
OUTPUT port of the adapter. 3. Turn the 7-mm OUTPUT port connector on the terminal adapter counterclockwise to secure it to the connector on the lower side of the test fixture. Figure 3-5 Connecting Test Fixture Keysight 42942A Terminal Adapter...
To compensate for any residual impedance and admittance existing in the test fixture, perform fixture compensation. The fixture compensation requires measurements with the 42942A for open and short compensation data. Generally, there is no need to perform load compensation. For more information on fixture compensation, refer to the Impedance Analyzer Help Manual.
3. Press the Load ( ) key to start the LOAD compensation data measurement. ( ) should indicate defined Load value. When the LOAD compensation data measurement is completed, the softkey label Load [OFF] (if it is off) changes to Load [ON]. Keysight 42942A Terminal Adapter...
Operation Fixture Extension Fixture Extension The 42942A is designed to connect directly to the Impedance Analyzer. Thus, the measurement point on the four-terminal pair cannot be extended by using four extension cables as shown in Figure 3-6 (A). The 42942A must be directly...
Specifications Specifications 4 Specifications This chapter provides specifications of the 42942A test fixture. Specifications Applicable Instruments Impedance Analyzer Frequency 20 Hz to 120 MHz Maximum Vol tage ±42 V peak max. (AC+DC) Operating temp. 0 °C to +40 °C Environment humid ity 15% to 95% RH ( @ wet bulb temp.
Service Maintenance 5 Service This chapter provides information on servicing and proper maintenance. Maintenance An exploded view of the 42942A of RoHS and non-RoHS compliance for parts identification is shown in Figure 5-1 Figure 5-2. Due to limited availability of RoHS compliance station and technical difficulties in RoHS soldering, only parts and support level that do not require RoHS soldering are supported.
Page 26
Service Maintenance Figure 5-1 Replaceable Parts (part 1 of 2) Keysight 42942A Terminal Adapter...