13 Maintaining the NPD

To Install a Capillary Column in the NPD

To Install a Capillary Column in the NPD

1 Gather the following materials:
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
WAR N IN G
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
WAR N IN G
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
3 Place a septum (if the column id is ≤0.1 mm), capillary column nut, and ferrule on the
Ferrule
Column nut
Septum
4 Score the column using a glass scribing tool. The score must be square to ensure a clean
Maintaining Your GC
Column
Ferrule(s), see
"Consumables and Parts for the NPD"
Column nut
Column cutter
1/4-inch open-end wrench
Septum
Isopropanol
Lab tissue
Lint-free gloves
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
column.
break.
on page 210.
213