Recommended separation distances between portable and mobile RF communications equipment and the ME equipment
The ME equipment is intended for use in an electromagnetic in which radiated RF disturbances are controlled. The customer or the user
of the ME equipment can help prevent electromagnetic interence by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the ME equipment as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output
power of transmitter
W
150 kHz to 80 MHz d=1,2
0.01
0.1
1
10
100
The ME system is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Vapor Expert(PY-001) Portable Nebulizer - Instruction Manual
Separation distance according to frequency of transmitter m
80 MHz to 800 MHz d=1,2
P
0.12
0.38
1.2
3.8
12
The ME declaration – electromagnetic immunity
Electromagnetic environment - guidance
3V
Portable and mobile RF communications equipment should be used no
closer to any part of the EQUIPMENT or SYSTEM including cables, than
the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.Interference may occur in
the vicinity of equipment marked with the following symbol.
3V/m
800 MHz to 2.5 GHz d=2,3
P
0.12
0.38
1.2
3.8
12
19
Mobi Mesh
/ English
P
0.23
0.73
2.3
7.3
23